GE Medical Systems
Information Technologies
Revision F
Servicing Manual Dash Responder V1.0
2004768-001
Page 31 of 72
•
Put the receptacle through the hole in the upper case shell. Make sure that the orientation of the
receptacle is right. The flex flat cable has to protrude out of the rubber part of the receptacle on the
side located to the back of the upper case shell. Regard the notch in the joint between the rubber
part of the receptacle and the (outer) plastic part. This notch has to be located to the back side of
the upper case shell. Fix the monitor connector receptacle with the screws.
•
The flat flex cable should be placed in the socket on the Power Management PCB before
reassembling the Power Management PCB to the device.
Make sure that the flex flat cable is inserted completely and perpendicular to the socket. Secure it
by fasten the actuator.
•
Carefully push the battery case into its place in the upper case shell. Make sure the rubber seal
around the opening of the battery case gets located in the frame properly. Be careful that the flex
flat cable doesn’t get cracked.