GE Medical Systems
Information Technologies
Revision F
Servicing Manual Dash Responder V1.0
2004768-001
Page 23 of 72
5
Service Menu
The S
ERVICE
M
ENU
can be reached by pressing the E
NERGY
S
ELECT
keys simultaneously for at least
two seconds.
User configurable settings
Some settings of the Dash Responder are configurable to meet the user’s requirements. The settings
can be made in the Service Menu and comprise following values:
•
Volume of the alarm sound
S
ERVICE
M
ODE
/ D
EVICE
S
ETUP
/ A
LARM
<L
OW
| M
EDIUM
| H
IGH
>
•
Automatic shutdown
S
ERVICE
M
ODE
/ D
EVICE
S
ETUP
/ S
HUTD
<O
N
| O
FF
>
•
User Language
S
ERVICE
M
ODE
/ D
EVICE
S
ETUP
/ L
ANGUAGE
< E
NGLISH
| D
EUTSCH
| … >
•
Autosequence of defib shocks
S
ERVICE
M
ODE
/ D
EFIB
S
ETUP
/ L
EVEL
1 <150 J | 200J | 300 J | 360 J>
S
ERVICE
M
ODE
/ D
EFIB
S
ETUP
/ L
EVEL
2 <150 J | 200J | 300 J | 360 J>
S
ERVICE
M
ODE
/ D
EFIB
S
ETUP
/ L
EVEL
3 <150 J | 200J | 300 J | 360 J>
•
Pacer Rate
S
ERVICE
M
ODE
/ P
ACER
S
ETUP
/ R
ATE
< 30 … 200
PPM
>
Component Test
Service Mode / Comp Test
Keyboard Test
S
ERVICE
M
ODE
/ C
OMP
T
EST
/ K
EY
T
EST
Press the keys for at least one second
Structure of the Service Menu
Main Service Menu
1
D
EVICE
S
ETUP
2
D
EFIB
S
ETUP
3
P
ACER
S
ETUP
4
S
ERVICE
M
ODE
5
E
XIT
S
AVE
&
E
XIT
Device Setup
2
M
AIN
M
ENU
1
D
ISPLAY
S
ETUP
6
A
LARM
<L
OW
>
S
HUTD
<O
FF
>
L
ANG
UAGE
7
F
ACT
S
ETT
Содержание Dash Responder
Страница 1: ...Dash Responder Version V 1 0 Servicing Instructions 2004768 001 ENG Revision F ...
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 84: ......
Страница 85: ......
Страница 91: ......
Страница 94: ......
Страница 95: ......
Страница 97: ......
Страница 98: ......
Страница 99: ......