APPENDIX A
13
Figure 3-1. Test Equipment Setup
d. Adjust the micrometer to a 15 mil (0. 38 mm) indication, then back to a 20 mil (0. 51 mm)
indication and record the output voltage. Going to the 15 mil (0. 38 mm) indication first will
eliminate any play in the micrometer before taking the 20 mil reading.
e. Increase the gap in 20 mil increments by adjusting the micrometer.
Record the voltage
indication at each increment.
f.
Either plot each voltage indication obtained in step e or compare each to the desired calibration
curve in Appendix C.
g. If the transducer is out of tolerance (refer to the
specifications),
recalibrate.
Содержание Bently Nevada 7200 Series
Страница 2: ...Part Number 74756 01 Rev G 07 07 Installation Manual 14 mm Proximity Transducer System 7200 Series ...
Страница 16: ...Section 2 INSTALLATION 11 Figure 2 4 Field Wiring and Shield Grounding for 7000 7200 and 9000 Series Monitors ...
Страница 22: ...APPENDIX A 17 Figure A 1 14 mm Probe Ordering Information ...
Страница 24: ...APPENDIX B 19 Figure A 3 14mm Extension Cable Ordering Information ...
Страница 34: ...APPENDIX C 29 6 APPENDIX C SCALE FACTORS AND SENSITIVITY CURVES ...
Страница 35: ...14 mm Proximity Transducer System 7200 Series Installation Manual 30 Figure C 1 Incremental Scale Factor ...
Страница 36: ...APPENDIX C 31 Figure C 2 Probe and One Foot Cable at Temperature ...
Страница 38: ...APPENDIX C 33 Figure C 4 9 Metres Cable at Temperature ...
Страница 42: ...APPENDIX C 37 Figure C 8 Effects of Flat Surface Side Clearance ...
Страница 43: ...14 mm Proximity Transducer System 7200 Series Installation Manual 38 Figure C 9 Effects of Counterbore ...