![GE Bently Nevada 7200 Series Скачать руководство пользователя страница 17](http://html.mh-extra.com/html/ge/bently-nevada-7200-series/bently-nevada-7200-series_installation-manual_132817017.webp)
14 mm Proximity Transducer System 7200 Series Installation Manual
12
3. MAINTENANCE
3.1 INTRODUCTION
This section contains calibration and troubleshooting procedures. The recommended maintenance
equipment is given in Table 3-1. If the recommended equipment is not available, equivalent
instruments can be used. Any maintenance performed by the user, other than that which is specified
herein, may void the warranty.
Table 3-1. Recommended Maintenance Equipment
MAINTENANCE EQUIPMENT
RECOMMENDED EQUIPMENT
Digital Multimeter
Test and Calibration Kit
Variable Resistor
Soldering Iron
Power Supply
Hewlett-Packard Model 3465 A/B
Bently Nevada Corporation Model TK3-2E or
TK3-2G
0 to 10k ohms
Weller Model SP23
Hewlett-Packard Model 6215A
3.2 CALIBRATION CHECK
Successful completion of the following procedure assures proper transducer calibration.
a. Connect test equipment as shown in Figure 3-1.
b. Adjust the spindle micrometer on the TK3 test and calibration kit until it indicates 20 mils (0.51
mm).
c. Insert the probe into the TK3 probe holder. Adjust the probe in the holder until the digital
multimeter indicates -2.00 ±0.01 Vdc and tighten the probe in the fixture. Do not contact the
probe tip.
Содержание Bently Nevada 7200 Series
Страница 2: ...Part Number 74756 01 Rev G 07 07 Installation Manual 14 mm Proximity Transducer System 7200 Series ...
Страница 16: ...Section 2 INSTALLATION 11 Figure 2 4 Field Wiring and Shield Grounding for 7000 7200 and 9000 Series Monitors ...
Страница 22: ...APPENDIX A 17 Figure A 1 14 mm Probe Ordering Information ...
Страница 24: ...APPENDIX B 19 Figure A 3 14mm Extension Cable Ordering Information ...
Страница 34: ...APPENDIX C 29 6 APPENDIX C SCALE FACTORS AND SENSITIVITY CURVES ...
Страница 35: ...14 mm Proximity Transducer System 7200 Series Installation Manual 30 Figure C 1 Incremental Scale Factor ...
Страница 36: ...APPENDIX C 31 Figure C 2 Probe and One Foot Cable at Temperature ...
Страница 38: ...APPENDIX C 33 Figure C 4 9 Metres Cable at Temperature ...
Страница 42: ...APPENDIX C 37 Figure C 8 Effects of Flat Surface Side Clearance ...
Страница 43: ...14 mm Proximity Transducer System 7200 Series Installation Manual 38 Figure C 9 Effects of Counterbore ...