PanaFlow™ LC User’s Manual
67
Chapter 3. Programming
fixture on the pipe (see Figure 74). When exiting from the
[Channel x]
menu, the meter will display a
message indicating that the Physical spacing of the transducers need to be adjusted to the value
calculated by the XMT1000.
4.
Adjust the physical spacing to value calculated by the XMT1000.
5.
Repeat section “Channel X Programming” on page 65 for all channels.
6.
This completes the programming for flow measurements. All further steps involve calibrating meter
for accurate velocity and sound speed measurements. Exit programming by pressing
[ESC]
until Save
options are displayed on the menu. Highlight
[Save]
or
[Save & Logout]
and Press
[ENTER]
to save
settings. The meter will not use the changed setting to make measurements until the settings are
explicitly saved.
7.
Make sure your pipe is full and there is no flow. Before you proceed to next section “Programming
Advanced Channel Settings” on page 69, allow for a 5-minute zero flow stabilization time.
Note:
If your transducers need to be installed with a spacing different than that calculated by the XMT1000,
make sure it is within 10% of the calculated value. Overwrite the meter calculated [Spacing] value with
installed Physical spacing.
Figure 73: Traverse Configurations
Содержание Baker Hughes PanaFlow LC XMT1000
Страница 1: ...Flow 910 327 Rev A Sep 2019 bhge com PanaFlow LC User s Manual ...
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 10: ...Contents viii PanaFlow LC User s Manual ...
Страница 14: ...Preface xii PanaFlow LC User s Manual no content intended for this page ...
Страница 104: ...Chapter 4 Error Codes and Troubleshooting 90 PanaFlow LC User s Manual no content intended for this page ...
Страница 106: ...Chapter 5 Maintenance and Service 92 PanaFlow LC User s Manual no content intended for this page ...
Страница 112: ...Appendix A Specifications and Model Configurations 98 PanaFlow LC User s Manual no content intended for this page ...
Страница 139: ...PanaFlow LC User s Manual 125 Appendix D Modbus Map Appendix D Modbus Map D 1 Input Registers Map ...
Страница 142: ...Appendix E CE Mark Compliance 128 PanaFlow LC User s Manual no content intended for this page ...
Страница 180: ...Appendix H Foundation Fieldbus Communication 166 PanaFlow LC User s Manual no content intended for this page ...
Страница 182: ...Warranty 168 PanaFlow LC User s Manual no content intended for this page ...
Страница 189: ......