Certification & Safety Statements for PanaFlow PF10-Zx
Apr 2018
CSS-0017, Rev. B
170 of 174
•
The product is an electrical apparatus and must be installed in the hazardous area in accordance with the
requirements of the EU Type Examination Certificate. The installation must be carried out in accordance with
all the appropriate international, national and local standard codes and practices and site regulations for
flameproof apparatus and in accordance with the instructions contained in the manual. Access to the
circuitry must not be made during operation.
•
The flameproof joints of the equipment are not intended to be repaired. Consult the manufacturer if
dimensional information on the flameproof joints is necessary.
•
Follow the manufacturer's instructions to reduce the potential of an electrostatic charging hazard.
•
Consult the manufacturer for genuine replacement flange fasteners. M10x35 hexagon socket cap screws of
ISO 12.9 DIN912 grade steel (zinc-plated) or better with a minimum yield strength of 135,000 psi are
acceptable alternatives.
•
Consult the manufacturer for genuine replacement flange fasteners. M10x35 hexagon socket cap screws of
ISO 12.9 DIN912 grade steel (zinc-plated) or better with a minimum yield strength of 135,000 psi are
acceptable alternatives.
Standards
The equipment complies with the standards listed in the following table:
ATEX/IEC
Specific Conditions of Use:
1
The flameproof joints of the equipment are not intended to be repaired. Consult the manufacturer if
dimensional information on the flameproof joints is necessary.
2
Follow the manufacturer's instructions to reduce the potential of an electrostatic charging hazard.
3
Consult the manufacturer for genuine replacement flange fasteners. M10x35 hexagon socket cap screws of
ISO 12.9 DIN912 grade steel (zinc-plated) or better with a minimum yield strength of 135,000 psi are
acceptable alternatives. M10x1.5, 365 mm Long, Zinc Plated, ISO 12.9 DIN912, 17-4 H1025 or better with
minimum yield strength of 145 KSI.
4
Consult the manufacturer for genuine replacement enclosure/adapter fasteners.
5
Care should be taken to avoid creating an ignition hazard due to impact or friction on the titanium
transducers which form part of the flameproof enclosure.
6
The electronics enclosure is rated for an ambient temperature range of -40°C to +60°C. The remote mount
junction box and flow body is rated for an ambient temperature range of -40°C to +60°C.
7
The equipment temperature class is dependent on the maximum process temperature and is according to
the following table:
Standards
IEC 60079-0:2011
EN 60079-0:2012 + A11:2013
IEC 60079-1:2014
EN 60079-1:2014
IEC 60529:2001
EN 60529:1991 + A1:2000
Содержание Baker Hughes PanaFlow LC XMT1000
Страница 1: ...Flow 910 327 Rev A Sep 2019 bhge com PanaFlow LC User s Manual ...
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 10: ...Contents viii PanaFlow LC User s Manual ...
Страница 14: ...Preface xii PanaFlow LC User s Manual no content intended for this page ...
Страница 104: ...Chapter 4 Error Codes and Troubleshooting 90 PanaFlow LC User s Manual no content intended for this page ...
Страница 106: ...Chapter 5 Maintenance and Service 92 PanaFlow LC User s Manual no content intended for this page ...
Страница 112: ...Appendix A Specifications and Model Configurations 98 PanaFlow LC User s Manual no content intended for this page ...
Страница 139: ...PanaFlow LC User s Manual 125 Appendix D Modbus Map Appendix D Modbus Map D 1 Input Registers Map ...
Страница 142: ...Appendix E CE Mark Compliance 128 PanaFlow LC User s Manual no content intended for this page ...
Страница 180: ...Appendix H Foundation Fieldbus Communication 166 PanaFlow LC User s Manual no content intended for this page ...
Страница 182: ...Warranty 168 PanaFlow LC User s Manual no content intended for this page ...
Страница 189: ......