PanaFlow™ LC User’s Manual
93
Appendix A. Specifications and Model Configurations
Appendix A. Specifications and Model Configurations
A.1 Operation and Performance
Fluid Types
Liquids: acoustically conductive fluids, including most clean liquids, and many liquids with small amounts
of entrained solids or gas bubbles. Maximum void fraction depends on transducer, interrogation carrier
frequency, path length, and pipe configuration.
Flow Measurement
Correlation transit time technique
Accuracy
•
±1%: >=2 in pipe; greater than 1 ft/s velocity
•
±2%: <=2 in pipe; greater than 1 ft/s velocity
Accuracy statement assumes measurement of a single phase homogenous liquid with a fully developed
symmetrical flow profile passing through the meter (typically 10 diameters upstream and 5 diameters
downstream of straight pipe run). Applications with piping arrangements (e.g. two out of plane elbow at
upstream) that create an asymmetrical flow profile may require extended piping straight runs and/ or flow
conditioning for the meter to perform to this specification.
Calibration
All meters are water calibrated and include a calibration certificate.
Repeatability
±0.15% of reading
Range (Bidirectional)
-82 to 82 ft/s (-25 to 25 m/s)
Rangeability (Overall)
50:1
Transducers supported
•
UTXDR 4 and 2 MHz
•
CF-LP 4 and 2 MHz
•
CRS 0.5, 1.0 and 2.0 MHz
•
CPT HT 0.5, 1.0 and 2.0 MHz
•
CET
Содержание Baker Hughes PanaFlow LC XMT1000
Страница 1: ...Flow 910 327 Rev A Sep 2019 bhge com PanaFlow LC User s Manual ...
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 10: ...Contents viii PanaFlow LC User s Manual ...
Страница 14: ...Preface xii PanaFlow LC User s Manual no content intended for this page ...
Страница 104: ...Chapter 4 Error Codes and Troubleshooting 90 PanaFlow LC User s Manual no content intended for this page ...
Страница 106: ...Chapter 5 Maintenance and Service 92 PanaFlow LC User s Manual no content intended for this page ...
Страница 112: ...Appendix A Specifications and Model Configurations 98 PanaFlow LC User s Manual no content intended for this page ...
Страница 139: ...PanaFlow LC User s Manual 125 Appendix D Modbus Map Appendix D Modbus Map D 1 Input Registers Map ...
Страница 142: ...Appendix E CE Mark Compliance 128 PanaFlow LC User s Manual no content intended for this page ...
Страница 180: ...Appendix H Foundation Fieldbus Communication 166 PanaFlow LC User s Manual no content intended for this page ...
Страница 182: ...Warranty 168 PanaFlow LC User s Manual no content intended for this page ...
Страница 189: ......