Chapter 3 Installation
3-39
3-9-3 Loading the System Software with USB Memory Stick (For R2.2.2 or above)
(cont’d)
6.) The system software load instruction window displays. Select one of the options for loading
the system. Select choice [a] to load the complete disk. Refer to
Figure 3-37 on page 3-39
-
To select [a], the complete disk will be loaded. This option is recommended for application
software upgrade.
NOTE:
When to select [a] to load complete disk, please ensure that any patient data on the disk has
been backed up.
-
To select [b], only the bootable C: partition is loaded. This option is intended for recovery of a
system that will not boot up. It is not recommended for application software upgrade because
during upgrade process, the data on the system would possibly be impacted.
-
To select [c] to quit system upgrade process.
NOTICE
If you select [a], ALL existing software and data will be erased. If backup has not been
performed, all data like Patient Database, System Configuration and User Configurations
(Customer Presets) will be lost.
Figure 3-37 System Software load instruction
WARNING
WARNING
While the software install procedure is designed to preserve data, you should select
choice [b] to format disk C only.