![GE HEALTHCARE Logiq Book XP series Скачать руководство пользователя страница 89](http://html.mh-extra.com/html/ge-healthcare/logiq-book-xp-series/logiq-book-xp-series_basic-service-manual_571895089.webp)
GE H
EALTHCARE
D
IRECTION
5194296-100, R
EVISION
10
LOGIQ B
OOK
XP S
ERIES
B
ASIC
S
ERVICE
M
ANUAL
Chapter 3 Installation
3-33
3-9-2
Loading the system software with DVD (For R2.2.2)
NOTE:
Because any data may be lost or impacted during upgrade process, the preserved patient data
on the system should be backed up before application software upgrade.
1.) Disconnect the system from the network.
2.) Power off the system.
3.) Connect the DVD-ROM to the system and turn on the DVD-ROM.
4.) Place the “Application Software” DVD into the DVD-ROM drive.
5.) Power on the system. The system will operate automatically to start application software
upgrading process.
6.) The system operates automatically to scan USB Devices as shown in
Figure 3-29 on page 3-33
WARNING
WARNING
While the software install procedure is designed to preserve data, you should save any
patient data, images, system setups to CD, DVD, USB Flash Drive, or USB Hard Disk.
Figure 3-29 Software Loading
Figure 3-30 Scanning USB Devices