![GE HEALTHCARE Logiq Book XP series Скачать руководство пользователя страница 176](http://html.mh-extra.com/html/ge-healthcare/logiq-book-xp-series/logiq-book-xp-series_basic-service-manual_571895176.webp)
GE H
EALTHCARE
D
IRECTION
5194296-100, R
EVISION
10
LOGIQ B
OOK
XP S
ERIES
B
ASIC
S
ERVICE
M
ANUAL
7-20
Section 7-6 - LOGIQ Book XP Series Diagnostic Descriptions
7-6-2
FRU Test
7-6-2-1
MST
These programs are provided for testing the MST board.
Table 7-3
FRU Test, MST Diagnostics Menu
Menu
Descriptions
Error Code
Overall Test
Performs all of the MST board tests listed below.
Assy Revision Test
10001
TRIG Test
The PC sends the command to the SH4 to generate the trigger for self test. Then the
results are returned to the PC.
10002
SH-4 Memory Test
The PC sends the command to the SH4 to check the memory especially for SH-4.
10003
SH-PCI Test 1
The PC sends the command to the SH4 to check the SDRAM in MST.
10004
SH-4 Test
Test the watchdog of SH4.
10005
COMSO BM Test
Use test mode of OQ card to check COMSO.
10006
Transmit HV Test
The HV unit outputs Plus and Minus THV, which is adjustable (variable).
This test checks if HVs can be read at A/D and adjusted by SH4.
10007
Supply HV Test
The HV unit generates Plus and Minus SHV for HVSW, which is adjusted to follow
THV.
Checks if it can be read at A/D in MST.
10008
LV Test
The LV unit outputs four kinds of DC Voltage.
This test checks if they can be read at A/D and D/A in MST.
10009
D/A-A/D Test
D/A-A/D channel test.
10010
USCBus Test
Check USCBus
Checks if the SH4 writes the data properly to USC Register or PMX Register.
10011
PCI/F1 Test (EUSC)
N/A
10012
PCI/F2 Test (IMGD)
N/A
10013
Memory Test
N/A
10014