GCE ECOlite4000 Скачать руководство пользователя страница 7

7/12

6.3.4 BLOCKING FUNCTION

Key Lock

To prevent buttons from being unintentionally pressed, a key lock can be activated by pressing the 

control b and – simultaneously. To deactivate it, press the control b and – again.

Locked flow

To change flow blocking:

• 

On the ECOlite

®

 4000 device that is switched off press the and hold it down. Afterwards, 

press the ON/OFF button and also hold it down. A PRESS KEY sign will appear.

• 

Press the two buttons and – simultaneously; a LANGUAGE sign will appear. 

• 

Using the ON/OFF button go to the BLOCKED setting, and using the +/- buttons set the mode 

(NO/YES). 

• 

You will confirm your selection and simultaneously exit the setting by pressing the ON/OFF button 

and holding it for approximately 2 seconds. 

• 

With activated fl ow blocking, the unit will be working with the fl ow rate set on the device before 

blocking was activated.

6.3.5 FUNCTIONAL TEST

• 

Switch on the ECOlite

®

 4000. 

• 

Choose a random setting, e.g. 3 l/m. 

• 

Breathe in through the cannula. 

• 

Make sure that there is a short puff of oxygen through the cannula by listening or feeling the flow. 

• 

Also make sure you can hear a slight click on each inhalation. 

• 

Check the display by pressing the “+” and “-” buttons; the setting should change accordingly.

• 

Activate the alarm by not breathing in the cannula. After approx. 1 minute the alarm should set 

off, both by sound and an alert on the display.

If there is an abnormal sound, noise, gas leakage or any other problem, stop using the 

product immediately and contact the supplier.
We recommend conducting a functional test (Chapter 6.3.5) every time before putting the 

device into operation.

6.3.6 ALARM AND MONITORING FUNCTIONS

A built-in microprocessor in the ECOlite

®

 4000 allows permanent monitoring of the most important 

functions:

Function 

Description 

Display screen

Audible 

alarm

Action

Normal  opera-

tion 

Indicates that the sen-

sor detects the start of 

an inhalation effort.

A picture of flash

-

ing light in top right 

hand corner

No

None

Low Oxygen 

supply

The oxygen flow 

from the source is 

insufficient to supply 

the patient. 

‘No O2’

Yes, after 

every 

exhalation

When this happens check:
The condition and tightness of 

spiral hose connections
For adequate oxygen supply
For any kinking and damage of 

the hose or the device.

Monitoring 

breath 

impulses

When the device is in 

use and no breathing 

takes place for one 

minute.

’CHECK 

CANNULA’

An audible 

signal is set 

off for 30 s, 

and the unit 

will then turn 

off automati-

cally.

The ECOlite

®

 4000 will switch off 

automatically, storing the current 

settings.
Check the cannula for 

obstructions and replace it if 

necessary.
(Remember to shut off oxygen 

supply – cylinder if the unit 

switches off.)

Содержание ECOlite4000

Страница 1: ...EN Page 1 12 ECOlite 4000 GAS SAVER INSTRUCTION FOR USE...

Страница 2: ...om the oxygen The dosage and timing of oxygen delivery is controlled by a microprocessor in accordance with the chosen setting Dosage per breath is independent of breathing rate therefore additional o...

Страница 3: ...TIONS At atmospheric pressure up to 60 oxygen is harmless 100 oxygen produces a cough after 6 hours and increasing breathlessness after 25 hours PERSONNEL INSTRUCTIONS AND TRAINING According to Medica...

Страница 4: ...function from national standards In the event of any problem with connection please contact the distributor the manufacturer Display screen ON OFF button Oxygen inlet Outlet into cannula Control butt...

Страница 5: ...clean If the unit is being used with an oxygen supply source with adjustable fl ow rates set the flow rate on the supply source to 4 litres per minute or to the closest setting above Check the cylinde...

Страница 6: ...s the button and hold it down Afterwards press the ON OFF button and also hold it down A PRESS KEY sign will appear Press the two buttons marked and simultaneously a LANGUAGE sign will appear Using th...

Страница 7: ...ngly Activate the alarm by not breathing in the cannula After approx 1 minute the alarm should set off both by sound and an alert on the display If there is an abnormal sound noise gas leakage or any...

Страница 8: ...void wasting oxygen 6 4 2 DISCONNECTING THE DEVICE Before disconnecting the ECOlite 4000 from the gas source Close the oxygen supply from the cylinder reservoir disconnect the connector of the spiral...

Страница 9: ...At the end of the product s life time 10 years at most the ECOlite 4000 device must be withdrawn from service To prevent reuse the owner shall make the ECOlite 4000 impossible to use or return it to...

Страница 10: ...c Area EEA This mark on the product indicates compliance with Directive 93 42 EEC relating to medical devices Class IIa Functional performance Sensitivity 0 13 cm of H2O Setting Manual automatic Trigg...

Страница 11: ...eter Use by Outlet parameter Take back equipment for recycling Do not dispose equipment into unsorted municipal waste WARRANTY The Standard Warranty period is two years from date of receipt by the GCE...

Страница 12: ...Asia The company operates 15 subsidiaries around the world and employs more than 850 people GCE Group includes four business areas Cutting Welding Process Applications Medical and High Purity Today s...

Отзывы: