Close the battery compartment cover.

Battery compartment cover

Battery compartment

Batteries LR6 1.5V 

Alkaline

• 

Fit the spiral hose onto the connector for oxygen inlet into the ECOlite

®

 4000 device by means 

of the square two-button adapter. A click should be felt when it is connected correctly. Pull the 

hose gently to confirm connection.

• 

Attach the other end of spiral hose to the oxygen source (pressure-reducing valve) using a 

standard connection (quick connector). 

• 

Attach a standard nasal cannula to the patient outlet connector on the ECOlite

®

 4000 by firmly 

pushing the tube over the connector.

6.2 SWITCHING ON

• 

Prior to every use, always check that correct components are delivered or used and make sure 

that they are complete and clean. 

• 

If the unit is being used with an oxygen supply source with adjustable fl ow rates, set the  flow 

rate on the supply source to 4 litres per minute or to the closest setting above. • Check the 

cylinder pressure gauge or LOX system to ensure sufficient oxygen supply, in the range of 1,6-5 

bar. 

• 

Open the oxygen cylinder valve or LOX system as instructed by your oxygen supplier. • Put on 

the supply nasal cannula. 

• 

Switch on the ECOlite

®

 4000 by firmly pressing the on/off button on the front of the unit for 2 

seconds. The ECOlite

®

 4000 will automatically perform a short operating check (ca. 4 s).

• 

The unit is now ready for use, and will operate at the basic settings.

Do not use the device if the cylinder content gauge on the supply valve is in the red low 

pressure zone.
Make sure that there is sufficient gas in the cylinders to complete the treatment. 
Oxygen used for medical purposes is a drug; when used incorrectly it may cause side 

effects.

®

®

Содержание ECOlite4000

Страница 1: ...EN Page 1 12 ECOlite 4000 GAS SAVER INSTRUCTION FOR USE...

Страница 2: ...om the oxygen The dosage and timing of oxygen delivery is controlled by a microprocessor in accordance with the chosen setting Dosage per breath is independent of breathing rate therefore additional o...

Страница 3: ...TIONS At atmospheric pressure up to 60 oxygen is harmless 100 oxygen produces a cough after 6 hours and increasing breathlessness after 25 hours PERSONNEL INSTRUCTIONS AND TRAINING According to Medica...

Страница 4: ...function from national standards In the event of any problem with connection please contact the distributor the manufacturer Display screen ON OFF button Oxygen inlet Outlet into cannula Control butt...

Страница 5: ...clean If the unit is being used with an oxygen supply source with adjustable fl ow rates set the flow rate on the supply source to 4 litres per minute or to the closest setting above Check the cylinde...

Страница 6: ...s the button and hold it down Afterwards press the ON OFF button and also hold it down A PRESS KEY sign will appear Press the two buttons marked and simultaneously a LANGUAGE sign will appear Using th...

Страница 7: ...ngly Activate the alarm by not breathing in the cannula After approx 1 minute the alarm should set off both by sound and an alert on the display If there is an abnormal sound noise gas leakage or any...

Страница 8: ...void wasting oxygen 6 4 2 DISCONNECTING THE DEVICE Before disconnecting the ECOlite 4000 from the gas source Close the oxygen supply from the cylinder reservoir disconnect the connector of the spiral...

Страница 9: ...At the end of the product s life time 10 years at most the ECOlite 4000 device must be withdrawn from service To prevent reuse the owner shall make the ECOlite 4000 impossible to use or return it to...

Страница 10: ...c Area EEA This mark on the product indicates compliance with Directive 93 42 EEC relating to medical devices Class IIa Functional performance Sensitivity 0 13 cm of H2O Setting Manual automatic Trigg...

Страница 11: ...eter Use by Outlet parameter Take back equipment for recycling Do not dispose equipment into unsorted municipal waste WARRANTY The Standard Warranty period is two years from date of receipt by the GCE...

Страница 12: ...Asia The company operates 15 subsidiaries around the world and employs more than 850 people GCE Group includes four business areas Cutting Welding Process Applications Medical and High Purity Today s...

Отзывы: