6.3 OPERATION

6.3.1 OPERATING MODE

The ECOlite

®

 4000 unit can be adjusted for two different operating modes:

Manual

 (shown as MANUAL on the display screen)

• 

In this mode the device will be delivering oxygen at the volume set on the display on each 

inhalation. 

• 

The value on the display is fi xed and will remain so regardless of the user’s breathing rate.

Automatic 

(shown as AUTO on the display screen)

• 

In the automatic mode the operation of the unit is dependent on the user’s breathing rate: 

• 

If the breathing rate is less than 15 breaths per minute, oxygen being delivered will be dosed in 

the volume selected by the user and shown on the display screen. 

• 

If the breathing rate is higher than 15 breaths per minute, the oxygen fl ow rate will automatically 

increase in a linear way up to double the flow rate of oxygen pre-set on the screen but it will not 

exceed the maximum flow rate of 8 litres per minute.

The preset oxygen fl ow rate will be indicated in full symbols and digits on the screen, with the 

additional amount indicated as half symbols.

The flow rate and operating mode shall be prescribed by your physician so that it best suits 

your heath condition.
When using the device, keep the connecting elements and the cannula in such a position in 

order to prevent any danger to the patient’s health or life.

To change the operating mode:

• 

On the ECOlite

®

 4000 device that is switched off press the and hold it down. Afterwards, 

press the ON/OFF button and also hold it down. A PRESS KEY sign will appear.

• 

Press the two buttons and – simultaneously; a LANGUAGE sign will appear. •

• 

Using the ON/OFF button go to the MAN/AUTO setting, and using the +/- buttons set the mode. 

• 

You will confirm your selection and simultaneously exit the setting by pressing the ON/OFF button 

and holding it for approximately 2 seconds.

6.3.2 OXYGEN SUPPLY VOLUMES

• 

Oxygen flow rates can be adjusted to values in 

the range of 0.5-8 litres per minute, as displayed 

on the screen by means of numerals and a 

scale. 

• 

Flow rates are adjusted in the interval of 0.5 

litre per minute by pressing the control buttons 

or – on the front side of the device.

• 

Supply settings can be adjusted at any time 

during use unless the fl ow blocking or key lock 

functions are being used (Chapter 6.3.4). 

• 

The sensitivity of the sensor in the device is 0.13 

cm of water column; the accuracy of oxygen 

dosing for a flow rate of 0.5-1.5 l/m is +/-30%, 

for a flow rate of 2-8 l/m it is +/-15%.

Current

flow rate

Operating

mode

Graphic display

 of range

Battery level 

indicator

6.3.3 LANGUAGES

To set the language:

• 

On the ECOlite

®

 4000 device that is switched off press the and hold it down. Afterwards, 

press the ON/OFF button and also hold it down. A PRESS KEY sign will appear. Press the two 

buttons and – simultaneously; a LANGUAGE sign will appear. 

• 

Set the required language by pressing the +/- buttons 

• 

You will confirm your selection and simultaneously exit the setting by pressing the ON/OFF button 

and holding it for approximately 2 seconds. 

®

Содержание ECOlite4000

Страница 1: ...EN Page 1 12 ECOlite 4000 GAS SAVER INSTRUCTION FOR USE...

Страница 2: ...om the oxygen The dosage and timing of oxygen delivery is controlled by a microprocessor in accordance with the chosen setting Dosage per breath is independent of breathing rate therefore additional o...

Страница 3: ...TIONS At atmospheric pressure up to 60 oxygen is harmless 100 oxygen produces a cough after 6 hours and increasing breathlessness after 25 hours PERSONNEL INSTRUCTIONS AND TRAINING According to Medica...

Страница 4: ...function from national standards In the event of any problem with connection please contact the distributor the manufacturer Display screen ON OFF button Oxygen inlet Outlet into cannula Control butt...

Страница 5: ...clean If the unit is being used with an oxygen supply source with adjustable fl ow rates set the flow rate on the supply source to 4 litres per minute or to the closest setting above Check the cylinde...

Страница 6: ...s the button and hold it down Afterwards press the ON OFF button and also hold it down A PRESS KEY sign will appear Press the two buttons marked and simultaneously a LANGUAGE sign will appear Using th...

Страница 7: ...ngly Activate the alarm by not breathing in the cannula After approx 1 minute the alarm should set off both by sound and an alert on the display If there is an abnormal sound noise gas leakage or any...

Страница 8: ...void wasting oxygen 6 4 2 DISCONNECTING THE DEVICE Before disconnecting the ECOlite 4000 from the gas source Close the oxygen supply from the cylinder reservoir disconnect the connector of the spiral...

Страница 9: ...At the end of the product s life time 10 years at most the ECOlite 4000 device must be withdrawn from service To prevent reuse the owner shall make the ECOlite 4000 impossible to use or return it to...

Страница 10: ...c Area EEA This mark on the product indicates compliance with Directive 93 42 EEC relating to medical devices Class IIa Functional performance Sensitivity 0 13 cm of H2O Setting Manual automatic Trigg...

Страница 11: ...eter Use by Outlet parameter Take back equipment for recycling Do not dispose equipment into unsorted municipal waste WARRANTY The Standard Warranty period is two years from date of receipt by the GCE...

Страница 12: ...Asia The company operates 15 subsidiaries around the world and employs more than 850 people GCE Group includes four business areas Cutting Welding Process Applications Medical and High Purity Today s...

Отзывы: