OPERATIONAL, TRANSPORT AND STORAGE SAFETY 

REQUIREMENTS

The ECOlite

®

 4000 may only be used with approved accessories and in a mode  prescribed by 

the physician. Using it for a therapy other than oxygen therapy is not permitted. Connection 

of the device to the oxygen source must conform to applicable national standards.
Never use a lubricant or oils with the oxygen supply device since spontaneous ignition may 

occur. Do not put the ECOlite

®

 4000 into operation with the presence of open  flames, e.g. 

during smoking, and do not use humidifiers.
The ECOlite

®

 4000 must be used in accordance with the information for EMC given in this 

Instruction for use.
Transportable and mobile high-frequency equipment can have influence on the ECOlite

®

 

4000.
Various equipment can influence each other during operation. To mitigate the influence  it 

might be necessary, for example, to change the position of such equipment or to shield it. 
The product and the related equipment must be protected from mechanical damage and 

kept out of the reach of unauthorised persons.
KEEP THE PRODUCT AND ITS ASSOCIATED EQUIPMENT AWAY FROM: 

• 

heat sources (fire, cigarettes, ...)

• 

flammable materials, 

• 

oil or grease (especially be careful if hand cream is used), 

• 

water,

• 

dust. 

National laws, decrees and regulations in relation to the prevention of accidents due to pressure 

equipment and environmental protection must be observed.

OXYGEN CONTRA-INDICATIONS

At atmospheric pressure, up to 60% oxygen is harmless. 100% oxygen produces a cough 

after 6 hours and increasing breathlessness after 25 hours.

PERSONNEL INSTRUCTIONS AND TRAINING

According to Medical Devices Directive 93/42/EEC the product provider must ensure that all 

personnel handling the product are provided with the operating instructions & performance data 

and are fully trained to carry out that operation. Trainees need to be supervised by an experienced 

person. 

Do not use the device without being trained. Training can be only done by a person with 

an appropriate education, experience and knowledge that has been also trained by the 

manufacturer.

For further information about training programs, please contact GCE.

PRODUCT DESCRIPTION

The ECOlite

®

 4000 is an electronic oxygen-dosing device with a built-in alarm function.

The ECOlite

®

 4000 has a built-in regulator that partially regulates the pressure. In order to work, 

it requires the supply of medical oxygen with a pressure of 1.6-5 bar and a minimum flow 4 litres 

per minute.

Oxygen flows from a cylinder (reservoir) through a pressure-reducing valve before entering  he 

ECOlite

®

 4000 device and then onto the patient via a nasal cannula. 

The incorporated microprocessor in the ECOlite

®

 4000 regulates and controls the oxygen flow to 

the patient.

The valve, which allows the oxygen to flow, is open only for a very short time until the next time 

when the patient inhales. This leads to the optimization of the amount of oxygen considering the 

course of inhalation and, thus, a considerable saving (by up to 2/3 of the volume), and mucous 

3

4

.

5

.

Содержание ECOlite4000

Страница 1: ...EN Page 1 12 ECOlite 4000 GAS SAVER INSTRUCTION FOR USE...

Страница 2: ...om the oxygen The dosage and timing of oxygen delivery is controlled by a microprocessor in accordance with the chosen setting Dosage per breath is independent of breathing rate therefore additional o...

Страница 3: ...TIONS At atmospheric pressure up to 60 oxygen is harmless 100 oxygen produces a cough after 6 hours and increasing breathlessness after 25 hours PERSONNEL INSTRUCTIONS AND TRAINING According to Medica...

Страница 4: ...function from national standards In the event of any problem with connection please contact the distributor the manufacturer Display screen ON OFF button Oxygen inlet Outlet into cannula Control butt...

Страница 5: ...clean If the unit is being used with an oxygen supply source with adjustable fl ow rates set the flow rate on the supply source to 4 litres per minute or to the closest setting above Check the cylinde...

Страница 6: ...s the button and hold it down Afterwards press the ON OFF button and also hold it down A PRESS KEY sign will appear Press the two buttons marked and simultaneously a LANGUAGE sign will appear Using th...

Страница 7: ...ngly Activate the alarm by not breathing in the cannula After approx 1 minute the alarm should set off both by sound and an alert on the display If there is an abnormal sound noise gas leakage or any...

Страница 8: ...void wasting oxygen 6 4 2 DISCONNECTING THE DEVICE Before disconnecting the ECOlite 4000 from the gas source Close the oxygen supply from the cylinder reservoir disconnect the connector of the spiral...

Страница 9: ...At the end of the product s life time 10 years at most the ECOlite 4000 device must be withdrawn from service To prevent reuse the owner shall make the ECOlite 4000 impossible to use or return it to...

Страница 10: ...c Area EEA This mark on the product indicates compliance with Directive 93 42 EEC relating to medical devices Class IIa Functional performance Sensitivity 0 13 cm of H2O Setting Manual automatic Trigg...

Страница 11: ...eter Use by Outlet parameter Take back equipment for recycling Do not dispose equipment into unsorted municipal waste WARRANTY The Standard Warranty period is two years from date of receipt by the GCE...

Страница 12: ...Asia The company operates 15 subsidiaries around the world and employs more than 850 people GCE Group includes four business areas Cutting Welding Process Applications Medical and High Purity Today s...

Отзывы: