FOREWORD

• 

The GCE ECOlite

®

 4000 unit is a medical device for conserving oxygen, classifi ed as Class IIa 

pursuant to

• 

Directive 93/42/EEC, concerning medical devices. 

• 

Compliance with the basic requirements of Directive 93/42/EEC is on the basis of EN ISO 

18779. 

• 

According to the type of protection from electrical shock, the product is included in Class BF.

• 

The product can be used simultaneously with a pacemaker.

INTENDED USE

ECOlite

®

 is for use by patients who have been prescribed oxygen therapy by a health professional. 

Oxygen therapy is intended for the treatment of a wide range of pulmonary diseases and support 

treatment for cardiovascular diseases.

The ECOlite

® 

equipment enables efficient dispensing of oxygen to the patient, by eliminating the 

wastage of oxygen which normally occurs with constant flow systems.

In principle a controlled pulse of oxygen is delivered to the patient in the initial phase of inspiration. 

This ensures the oxygen gas reaches the alveoli in the initial phase of inspiration and the remaining 

airways are not unnecessarily filled with oxygen. The oxygen saving is substantial and the mucous 

membrane is proportionally less stressed due to the reduced drying out effect from the oxygen.

The dosage and timing of oxygen delivery is controlled by a microprocessor in accordance with the 

chosen setting. Dosage per breath is independent of breathing rate, therefore additional oxygen is 

automatically delivered to the patient when breathing rate increases.

2.1 IDENTIFIED PATIENT POPULATION

• 

Age

 

 

not specified

• 

Weight

  

 

not specified

• 

Health status 

 

the user should not show signs of contra-indications associated with    

    

the application of medical oxygen

2.2 THE DESIGNATED BODY PART OR TISSUE TYPE , WHICH THE 

APPLICATION RELATES OR TO WHICH IT IS TO OPERATE

• 

Oxygen therapeutic substance passes into the airways of the lungs and acts through the 

cardiovascular system

• 

Cannula for application of the oxygen in contact with the skin of the face

2.3 INTENDED CONDITIONS OF USE

• 

Environment - designed for outdoor and indoor use (see conditions Chapter 9.5)

• 

Frequency of use: Designed for multiple uses, daily use , one time use is limited by battery 

capacity and oxygen bottles .

• 

Location : Protect from physical damage.

• 

Mobility: The device is portable, battery-powered .

2.4 INTENDED PROFILE

• 

The device is intended for home using

• 

Education:

  

are not subject to requirements for specialized education

• 

Knowledge:

  

training of doctors prescribing therapy , the study and knowledge of    

    

the user manual

• 

Skills:

 

 

Ability to operate electronic devices

2.5 OPERATING PRINCIPLES

• 

It must not come into contact with grease.

1

.

2

.

Содержание ECOlite4000

Страница 1: ...EN Page 1 12 ECOlite 4000 GAS SAVER INSTRUCTION FOR USE...

Страница 2: ...om the oxygen The dosage and timing of oxygen delivery is controlled by a microprocessor in accordance with the chosen setting Dosage per breath is independent of breathing rate therefore additional o...

Страница 3: ...TIONS At atmospheric pressure up to 60 oxygen is harmless 100 oxygen produces a cough after 6 hours and increasing breathlessness after 25 hours PERSONNEL INSTRUCTIONS AND TRAINING According to Medica...

Страница 4: ...function from national standards In the event of any problem with connection please contact the distributor the manufacturer Display screen ON OFF button Oxygen inlet Outlet into cannula Control butt...

Страница 5: ...clean If the unit is being used with an oxygen supply source with adjustable fl ow rates set the flow rate on the supply source to 4 litres per minute or to the closest setting above Check the cylinde...

Страница 6: ...s the button and hold it down Afterwards press the ON OFF button and also hold it down A PRESS KEY sign will appear Press the two buttons marked and simultaneously a LANGUAGE sign will appear Using th...

Страница 7: ...ngly Activate the alarm by not breathing in the cannula After approx 1 minute the alarm should set off both by sound and an alert on the display If there is an abnormal sound noise gas leakage or any...

Страница 8: ...void wasting oxygen 6 4 2 DISCONNECTING THE DEVICE Before disconnecting the ECOlite 4000 from the gas source Close the oxygen supply from the cylinder reservoir disconnect the connector of the spiral...

Страница 9: ...At the end of the product s life time 10 years at most the ECOlite 4000 device must be withdrawn from service To prevent reuse the owner shall make the ECOlite 4000 impossible to use or return it to...

Страница 10: ...c Area EEA This mark on the product indicates compliance with Directive 93 42 EEC relating to medical devices Class IIa Functional performance Sensitivity 0 13 cm of H2O Setting Manual automatic Trigg...

Страница 11: ...eter Use by Outlet parameter Take back equipment for recycling Do not dispose equipment into unsorted municipal waste WARRANTY The Standard Warranty period is two years from date of receipt by the GCE...

Страница 12: ...Asia The company operates 15 subsidiaries around the world and employs more than 850 people GCE Group includes four business areas Cutting Welding Process Applications Medical and High Purity Today s...

Отзывы: