Appendix B:  
Calibration 
Frequency 

One of the most common 
questions that we are asked at 
Sperian Instrumentation is: 

“How 

often should I calibrate my gas 
detector?”

 

Sensor Reliability and Accuracy 

Today’s sensors are designed to 
provide years of reliable service.  
In fact, many sensors are designed 
so that with normal use they will 
only lose 5% of their sensitivity per 
year or 10% over a two-year 
period.  Given this, it should be 
possible to use a sensor for up to 
two full years without significant 

loss of sensitivity. 

Verification of Accuracy 

With so many reasons why a 
sensor can lose sensitivity and 
given the fact that dependable 
sensors can be key to survival in a 
hazardous environment, frequent 
verification of sensor performance 

is paramount. 
There is only one sure way to 
verify that a sensor can respond to 
the gas for which it is designed. 
That is to expose it to a known 
concentration of target gas and 
compare the reading with the 
concentration of the gas.  This is 
referred to as a “bump” test. This 
test is very simple and takes only a 
few seconds to accomplish. 

The 

safest course of action is to do a 
“bump” test prior to each day’s 
use

. It is not necessary to make a 

calibration adjustment if the 
readings fall between 90%* and 
120% of the expected value. As an 
example, if a CO sensor is 
checked using a gas concentration 
of 50 PPM it is not necessary to 
perform a calibration unless the 
readings are either below 45 PPM 

or above 60 PPM. 

*The Canadian Standards 
Association (CSA) requires 
combustible gas sensors to 
undergo calibration when the 
displayed value during a bump 
test fails to fall between 100% 
and 120% of the expected value 

for the gas. 

Lengthening the Intervals 

between Verification of 

Accuracy 

We are often asked whether there 
are any circumstances in which the 
period between accuracy checks 

may be lengthened.  
Sperian Instrumentation is not the 
only manufacturer to be asked this 
question! One of the professional 
organizations to which Sperian 
Instrumentation belongs is the 
Industrial Safety Equipment 
Association (ISEA). The 
“Instrument Products” group of this 
organization has been very active 
in developing a protocol to clarify 
the minimum conditions under 
which the interval between 
accuracy checks may be 

lengthened. 
A number of leading gas detection 
equipment manufacturers have 
participated in the development of 
the ISEA guidelines concerning 
calibration frequency. Sperian 
Instrumentation’s procedures 

closely follow these guidelines.  
If your operating procedures do not 
permit daily checking of the 
sensors, Sperian Instrumentation 
recommends the following 
procedure to establish a safe and 
prudent accuracy check schedule 

for your Sperian instruments: 
1.  During a period of initial use of 

at least 10 days in the 
intended atmosphere, check 
the sensor response daily to 
be sure there is nothing in the 
atmosphere that is poisoning 
the sensor(s). The period of 
initial use must be of sufficient 
duration to ensure that the 
sensors are exposed to all 
conditions that might have an 

adverse effect on the sensors.  

2.  If these tests demonstrate that 

it is not necessary to make 
adjustments, the time between 
checks may be lengthened. 
The interval between accuracy 
checking should not exceed 

30 days.  

3. 

When the interval has been 
extended the toxic and 
combustible gas sensors 
should be replaced 
immediately upon warranty 
expiration. This will minimize 
the risk of failure during the 
interval between sensor 

checks. 

 

4.  The history of the instrument 

response between 
verifications should be kept. 
Any conditions, incidents, 
experiences, or exposure to 

contaminants that might have 
an adverse effect on the 
calibration state of the sensors 
should trigger immediate re-
verification of accuracy before 
further use.  

5.  Any changes in the 

environment in which the 
instrument is being used, or 
changes in the work that is 
being performed, should 
trigger a resumption of daily 

checking.  

6.  If there is any doubt at any 

time as to the accuracy of the 
sensors, verify the accuracy of 
the sensors by exposing them 
to known concentration test 
gas before further use.  

Gas detectors used for the 
detection of oxygen deficiencies, 
flammable gases and vapors, or 
toxic contaminants must be 
maintained and operated properly 
to do the job they were designed to 
do. Always follow the guidelines 
provided by the manufacturer for 
any gas detection equipment you 

use! 
If there is any doubt regarding your 
gas detector's accuracy, do an 
accuracy check! All it takes is a 
few moments to verify whether or 
not your instruments are safe to 

use. 

 

One Button Auto Calibration 

While it is only necessary to do a 
“bump” test to ensure that the 
sensors are working properly, all 
current Sperian gas detectors offer 
a one-button auto calibration 
feature. This feature allows you to 
calibrate a Sperian gas detector in 
about the same time as it takes to 
complete a “bump” test. The use of 
automatic bump test and 
calibration stations can further 
simplify the tasks, while 

automatically maintaining records. 

Don't take a chance  

with your life.   

Verify accuracy frequently! 

Please read also Sperian 
Instrumentation’s application note: 

AN20010808 “Use of ‘equivalent’ 
calibration gas mixtures”

.  This 

application note provides 
procedures to ensure safe 
calibration of LEL sensors that are 
subject to silicone poisoning.

 

Biosystems website is located at 

http://www.biosystems.com 

Содержание IQ Express

Страница 1: ...R eferenc e M anual IQ E xpres s Doc king S tation for ToxiPro ToxiLtd Toxi3Ltd Gastech 24 Baretta Rd Wangara Western Australia 6065 Tel 1800 999 902 Fax 1800 999 903 http www gastech com ...

Страница 2: ...NDERSTOOD AND FOLLOWED THIS MANUAL IS NOT INTENDED TO REPLACE THE OPERATIONS AND OR REFERENCE MANUALS FOR THE GAS DETECTOR THIS MANUAL IS ONLY DESIGNED TO AID IN THE INSTALLATION AND OPERATION OF THE IQ EXPRESS SYSTEM AND SHOULD BE USED IN CONJUNCTION WITH THE INSTRUMENT REFERENCE OR OPERATIONS MANUAL AT ALL TIMES IQ Express Reference Manual Sperian Instrumentation Part Number 13 275 Version 4 02 ...

Страница 3: ... Manager Software Installation 7 2 4 IQ Express PC Software Installation 7 2 5 Install IQ Administrator Pro and PostgreSQL database 8 2 6 Power and Connectivity 8 2 6 1 PC Single Dock via USB 8 2 6 2 PC Multi Dock via USB 8 2 7 Connecting the Calibration Gas Cylinder s to the Dock s 9 2 7 1 Single Cylinder to Single IQ Express Dock 9 2 7 2 Single Cylinder to Multiple Docks 10 2 7 3 Fresh Air Port ...

Страница 4: ...ates 24 5 4 Session Downloading 24 6 Software Revisions 24 Appendix A USB Architecture with Network Access 25 Appendix B Calibration Frequency 26 Sperian Instrumentation Warranty Gas Detection Products 27 Warnings and Cautions A Signal Words The following signal words as defined by ANSI Z535 4 1998 are used in the IQ Express Reference Manual indicates an imminently hazardous situation which if not...

Страница 5: ...d the PC has been restarted 1 1 Tests and record keeping The IQ Express Station automatically performs the following eight procedures whenever an instrument is placed in the dock Instrument identification Battery test Audible alarm test Visual alarm test Vibrating alarm test if applicable Fresh air calibration Bump test Record keeping Datalogger and Eventlogger download IQ Express must be connecte...

Страница 6: ... or if the dock is to be connected via Ethernet the IQ Express must be plugged in to an appropriate electrical outlet using the power supply wall cube that was included with the dock at the time of purchase If multiple IQ Express docks are purchased with the IQ Express Mounting Bracket an optional power supply is available that is capable of running up to four IQ Express Docks from a single electr...

Страница 7: ...xpress Docks are shipped from Sperian Instrumentation configured for CO H2S instruments with the standard calibration gas settings If the user chooses to activate the IQ Express without a PC connection or if the PC connection has been lost the dock will operate with the last set of calibration gas settings that were uploaded into the instrument and will automatically store calibration data until t...

Страница 8: ...Windows Explorer 2 Select IQ Express Pro and click Install The InstallShield Wizard will begin the installation 3 Click Next to continue with the installation The Licensing Agreement screen will be shown 4 If you accept the terms of the License Agreement click I accept and then click Next to continue The upgrade information and PC requirements will be shown 5 Click Next to continue The Customer In...

Страница 9: ...7 for calibration gas cylinder connection instructions If you have purchased the IQ Express Mounting Bracket proceed to the Assembly Instructions for IQ Express Mounting Bracket Once the docks are correctly installed on the mounting bracket proceed to section 2 7 3 for instructions concerning the Fresh Air Port s 2 6 1 PC Single Dock via USB For single dock setups the power from the PC via the USB...

Страница 10: ...onnected to the cylinder and whether the calibration gas is considered reactive Enhanced versions of the IQ Express include a tubing assembly that is suitable for use with reactive gases Figure 2 7 IQ Express Gas Ports For instructions on connecting the gas cylinder to a single IQ Express dock proceed to section 2 7 1 For instructions on connecting the gas cylinder to multiple IQ Express docks pro...

Страница 11: ...as while using the 12 039 demand flow regulator The 54 46 115 manifold is comprised of 1 small black elbow fitting 3 small black T fittings 2 pieces of gas tubing 5 long 1 piece of gas tubing 10 long 1 piece of gas tubing 1 long 54 46 0115 Manifold 1 Insert the Demand Flow Regulator into the calibration gas cylinder 2 Modify the tubing assembly as follows depending on how many docks will be connec...

Страница 12: ... dock The air filter should be installed if the dock is drawing fresh air from the immediate surroundings or if it is drawing fresh air from an external source via a sealed conduit The fresh air filter can be left off if the fresh air source is a calibration cylinder Only one side of the filter will screw into the Air Port on the dock Providing Fresh Air in a Contaminated Environment If the IQ Exp...

Страница 13: ...mply plumb the fresh air source directly into the dock s AIR port For multiple dock setups in which fresh air is provided to the dock via sealed conduit begin by installing the fresh air filter on each dock A one way check valve must be placed in the line between the t fitting in the fresh air manifold and each dock to ensure the purity of the fresh air source See Multi Dock Setups with Fresh Air ...

Страница 14: ...ew Hardware Wizard screen will be shown on the PC NOTE These screens vary depending on your PC s operating system Click Next to continue The Wizard will proceed to search for the USB drivers If installing to XP the following screen will be shown and should be disregarded Click Continue Anyway XP Only The software will notify you when the process is complete Click Finish The IQ Express Dock will di...

Страница 15: ... Dock status LED The status bars at the far left show the status of the docks that are currently registered on the PC Each dock that is currently connected to the PC by either USB or by a Live Ethernet connection will have a distinct status bar The green LED at the far left shows the general status of the dock All docks that are connected via Ethernet and are currently in Listening Mode are given ...

Страница 16: ...tly recognized in the dock The last tab on the left accesses all of the IQ Express Docks that are connected by Ethernet and in Listening Mode 3 2 1 Dock Configuration Click on the Dock tab to access the dock settings To change the configuration of the IQ Express Dock that is currently connected to the PC click on the Change Config button at the lower right of the screen Once the dock configuration...

Страница 17: ...ting is controlled by a slider The higher the setting the louder the alarm will need to be to pass the test The sound level is given in decibels when the cursor is placed over the slider Calibration Interval The Dock s calibration interval settings are controlled from the Dock Tab and are located below the battery level and above the LCD Display box The dock s calibration interval is the maximum n...

Страница 18: ...ate and potentially dangerous readings Special Instructions Gas Configuration Oxygen Detectors An oxygen detector can be processed by the IQ Express in a fresh air atmosphere without the use of calibration gas but an optional bump test is available to verify oxygen sensor response The O2 Bump and Calibration is automatically enabled at the factory If the user wishes to bypass the oxygen bump test ...

Страница 19: ...click on the Ethernet button To test the dock connection press the Ping button A status window will be shown to indicate whether the ping was successful or not The dock s e mail settings are controlled through the Database Manager Pro program but can be viewed by clicking on the e mail button The e mail status IP Address of the dock recipient e mail address and text of the e mail message are shown...

Страница 20: ...d set to check for the code in the instruments 1 Verify that the ToxiPro and ToxiLtd instruments have firmware version 6 3 or higher Instrument firmware version is shown in the start up screens for the instrument If needed new ToxiPro Firmware can be downloaded at no charge from Sperian Instrumentation s download website at http www biodownloads com The ToxiLtd will need to be returned to Sperian ...

Страница 21: ...ton Enter your User ID and password again when the Security Login window is shown A window will be shown to indicate that the dock configuration has been updated 11 Click OK PIN Lock Enabled will then be shown in the PC software in the dock configuration window The dock will now verify the PIN in any instrument that interfaces with it 12 For every other dock in the system access the dock s advance...

Страница 22: ...e if the instrument passes the bump test The gas detector will pass the bump test if the sensor response to the gas is within 10 if the reading is lower than the expected sensor output or within 20 if the reading is higher than the expected sensor output If the bump test is passed the IQ Express will proceed to the alarm tests If the instrument fails the bump test or if the instrument s calibratio...

Страница 23: ...ed during calibration In the event that a detector is placed in the dock that is incompatible with the IQ Express Dock s calibration gas setting the dock will display WRONG GAS TYPE and the interface will be stopped The output box in the software will indicate that the gas type is not supported the fault indicator will be lit and the test will not be performed Remove the instrument from the IQ Exp...

Страница 24: ...atically update ToxiPro Firmware and calibration options The Timeout setting controls the amount of time that needs to pass after connection between the dock and the PC is lost before the instrument will reset itself Options are in 5 second increments from 5 to 30 seconds To change the setting click on the arrow to the right of the timeout setting to access the list and select the new setting The ...

Страница 25: ... directory for the downloaded files either type the address in directly or use the Browse button to locate the new folder To view stored data files in the download folder click on the Show Data Files button Once the data files are shown if BioTrak version 7 40 or later is installed clicking on one of the stored files will automatically launch BioTrak 6 Software Revisions Sperian Instrumentation wi...

Страница 26: ...25 Appendix A USB Architecture with Network Access ...

Страница 27: ...mmends the following procedure to establish a safe and prudent accuracy check schedule for your Sperian instruments 1 During a period of initial use of at least 10 days in the intended atmosphere check the sensor response daily to be sure there is nothing in the atmosphere that is poisoning the sensor s The period of initial use must be of sufficient duration to ensure that the sensors are exposed...

Страница 28: ... purchase ToxiLtd 2 years after activation or 2 years after the Must Be Activated By date whichever comes first Toxi3Ltd 3 years after activation or 3 years after the Must Be Activated By date whichever comes first Mighty Tox 90 days after activation or 90 days after the Must Be Activated By date whichever comes first Mighty Tox 2 Prorated credit is given towards repair or purchase of a new unit o...

Отзывы: