F-0X function table
Note:
F-06 and below operations do not need to be operated under normal circumstances, and only
professionals can operate when necessary. F-11 to F-19 are reserved functions, please consult the
manufacturer for details.
F-01
address settings
RS485 communication address modification
F-02
low alarm value setting The setting value cannot exceed the high alarm value
F-03
High alarm value setting
F-04
Zero calibration
The setting value cannot be lower than the low alarm value
Calibration after stable reading in clean environment
F-05
4MA current trimming
C-04, 4mA output (modifiable)
C-08, 8mA output (cannot be modified)
C-12, 12mA output (cannot be modified)
C-16, 16mA output (cannot be modified)
C-20, 20mA output (modifiable)
F-06
ADC value
Cannot be modified
F-07
Set calibration point
Span calibration concentration value
F-08
Span calibration
Change 0 to 1, press the OK key, the real-time concentration value will
be displayed, and then press the OK key to calibrate as the value set in
F-07
F-09
Production date
cannot be modified
F-10
Communication
protocol
0:Standard version
1:Non-standard version (old protocol)