background image

4.

 At this time, pass in the calibration gas at 300±20ml/min. When 

the value of the instrument is stable, press the "OK" key. The instrument 

will be automatically calibrated and display "PASS" for a short time. At 

this time, the measured concentration value is consistent with the value 

set in "F-07" (concentration value of calibration gas).

IR AL AH

Note: 

Pass the calibration gas into the detector. After the output is 

stable, adjust and correct the display value of the detector to be the 

same as the calibration gas value, and then cut off the gas. Observe 

whether it can return to the zero point (in a pure air environment) or 
the starting point, and then repeat it again, and the calibration can be

 

IR  ALAH

completed with a small difference between the two values (within the 

basic error range). If there is a big difference, repeat the above method to calibrate until the calibration 

specification is met (the detector is usually adjusted to "20.93% VOL" for oxygen in the air environment, and 

the detector is usually adjusted to "78.10% VOL" for nitrogen in the air environment).

Remote controller key function

OK

BACK

MENU

RESET

6  Remote controller operation

1: Press the MENU key once to enter the F-01 address code setting menu, press the "OK" key to enter 

the modification, press the "+" and "-" keys to modify the instrument address code, and press the "OK" key to 

save after the modification is completed. Press the "BACK" key to cancel.

2: Press the "+" key to enter F-02 to enter the low alarm point setting menu, and press the "OK" key to 

enter the modification. Press the "+" or "-" key to modify the low alarm value of the instrument. After the 

modification is completed, press the "OK" key to save, and press the "BACK" key to cancel.

3: Continue to press the "+" key to enter F-03 to enter the high alarm point setting menu, and press the 

"OK" key to enter the modification. Press the "+" and "-" keys to modify the high alarm value of the 

instrument. After the modification is completed, press the "OK" key to save, and press the "BACK" key to 

cancel.

4: In the parameter setting state, if the remote controller is not operated for 10 seconds, the 

instrument will automatically return to the gas detection state, and the parameters being modified 

(without pressing the "OK" key to confirm the operation) will not be saved in the EEPROM. The 

parameters used  are still the previous values when restarted again.

Key

Function

MENU

  Enter the parameter setting state from the

detection state

RESET

  Cancel the alarm state

+

  Displayed parameter increase 1

-

  Displayed parameter minus 1

OK

  Confirm the operation

BACK

Cancel operation/return

13

www.gasdog.com

[email protected]

+86 773-280-9002

Содержание GD300 Series

Страница 1: ...Online Fixed Gas Detector GD300 XX Instruction Manual Notice to Reader Please read this instruction manual before using the detector 1 www gasdog com sales gasdog com 86 773 280 9002...

Страница 2: ...ery 3 6 months the explosion proof disc of the sensor should be cleaned regularly purging with low pressure compressed air according to the applicable scene otherwise dust and impurities will block th...

Страница 3: ...re operation 2 Structure diagram 3 Installation method 4 Wiring Instructions 5 Calibration method 6 Remote controller operation 7 Common faults and repairs 8 Instrument con guration 9 Appendix 3 www g...

Страница 4: ...many alarms or the long term alarm state will make people paralyzed if it is too far the reaction speed will be too slow or even no response B The instrument should be installed in the downwind direc...

Страница 5: ...ype The installation thread size of the instrument is M45 P2 0 During the installation process hold the detector with both hands so that the sensor is parallel to the installation thread align the scr...

Страница 6: ...rance and installation hole map 118 00 8 00 140 00 8 00 8 00 102 30 206 00 89 00 Apperance and installation hole map of gas detector with alarm light 8 00 179 00 102 00 8 00 8 00 118 00 206 00 89 00 6...

Страница 7: ...mula The maximum current of combustible gas detector is 250mA and the maximum loop resistance is 48 The maximum current of the toxic gas detector is 100mA and the maximum loop resistance is 120 The ca...

Страница 8: ...the load resistance which includes the output resistance of the control system controller DCS or PLC and the internal resistance of the cable Cable distance m 100 200 500 1000 The maximum allowable n...

Страница 9: ...needs to open the instrument shell and connect wires to the inside terminals The following gure is the description of each terminal Warning The wiring work must be performed by professionals otherwis...

Страница 10: ...the alarm settings of the detector and the panel are adjusted to the factory default values and the user can adjust it through the remote controller according to the situation Zero calibration Make su...

Страница 11: ...zero calibration is successful it will automatically return to the detection interface IR ALAH Span calibration Connect the calibration gas cylinder of known concentration to the calibration cap Opti...

Страница 12: ...i ed by the and keys to make the value consistent with the calibration gas concentration Press the OK key to display PASS brie y IR ALAH IR ALAH IR ALAH IR ALAH IR ALAH IR ALAH 2 Press the key to disp...

Страница 13: ...ting menu press the OK key to enter the modi cation press the and keys to modify the instrument address code and press the OK key to save after the modi cation is completed Press the BACK key to cance...

Страница 14: ...value Calibration after stable reading in clean environment F 05 4MA current trimming C 04 4mA output modi able C 08 8mA output cannot be modi ed C 12 12mA output cannot be modi ed C 16 16mA output c...

Страница 15: ...ot enough warm up time Sensor failure Circuit failure Send back for repair recalibrate Reset the detector range Replace the sensor recalibrate Reset the detector range Replace the sensor Power on and...

Страница 16: ...tively These two signal output modes are also di erent when wiring A 4 20mA current signal output connection mode must be grounded together B RS485 digital signal output connection mode 1 can be indep...