BreathSpec®
– Quickstart Manual
49/67
18
Switch on both devices.
19
Option
For file transfer via LAN:
Connect the
Ethernet
socket with an
appropriate Ethernet
cable to a LAN.
INFORMATION!
For detailed information about the installation and operation of the
µTD please refer the
µTD-GC-IMS Additional Information
Quickstart
and the
Airsense µTD Manual.
INFORMATION!
When the µTD unit is connected to BreathSpec®, the current loop
data output cannot be used.
6.3 Prepare the device for operation
Before using the device for the first time or after being disconnected from the
nitrogen source for some time it must be cleaned to ensure proper operating
conditions. In this case start the
cleaning mode
.
An appropriate cleaning period duration must be chosen depending on how long the
device was switched off and on the extent of contamination. It is recommended to
clean the device for at least
15 hours
before operating it for the first time. If the
Spectrum is not clean repeat the process. A
reference spectrum
for the evaluation
can be found in the supplied
Analytical Approval
.