BreathSpec®
– Quickstart Manual
47/67
The exhaust tubes (Gas out and Sample gas out)
must be led
separately into the exhaust system and must not be connected.
The exhaust system must
not generate any negative pressure
.
11
Connect TD operating
gas in socket to
Driftgas-
/ Carriergas-Adapter
using a 3 mm Swagelok
PFA tube.
12
Also connect Outlet
ZeroGas 2
and
TDZero
Gas 2
using a 3 mm
Swagelok PFA tube.
For further Information
see: Airsense µTD
manual
13
Connect BreathSpec
Signal Converter I/O-
Port
to
µTD Digital Port
with appropriate
Connection-Cable.