Instrument Records (Logs):  

During operation the instrument keeps an active log of all instrument activity. These records 
can be downloaded from the instrument via the MGC IR Link software and an IR Link, or 
from the Clip Dock. 

The Event, Bump, and Calibration Logs are always downloaded. The user may also choose 
to download different amounts of the data log to reduce the transfer time. Partial data logs 
contain approximately 1 weeks’ worth of data. New data log downloads will only contain the 
data since the last download. Full log downloads will download the entire data log.

 

Event Log 

The detector stores the last 25 alarm events. These are organized by first in first out (FIFO) 
so the 26

th

 event will replace the first event, and so on. The detector records the alarm 

conditions for each event: 

 

Date and time at start of event 

 

Duration of the alarm condition 

 

Each sensor’s peak alarm status and reading 

Bump Log 

The detector stores the last 25 alarm events. These are organized by first in first out (FIFO) 
so the 26

th

 event will replace the first event, and so on. Bump tests are differentiated from 

normal events when the alarm condition occurs inside of a Clip Dock test station, or when 
the instrument is manually bump tested according to the Bump Testing section. The 
detector records the bump status for each test: 

 

Date and time of bump test 

 

Whether the test was performed manually or in a Clip Dock 

 

Each sensor’s peak alarm status and reading 

 

The result of each sensor’s bump test 

Calibration Log 

The detector records the last 25 calibration attempts. These are organized by first in first 
out (FIFO) so the 26

th

 calibration will replace the first calibration and so on. Each calibration 

attempt will be recorded: 

 

Date and time of calibration 

 

Each sensor’s gas concentration calibrated to 

 

Each sensor’s calibration success status 

Data Logs 

The detector records its current operational status every second. The logging interval 
cannot be adjusted, but the detector compresses the data to reduce the storage and 
transfer times of redundant records. The following items are recorded into the log: 

 

Date and time 

 

Sensor readings and status conditions 

 

All user and sensor options 

 

Events (i.e. turn on or off) 

Содержание Multi Gas Clip

Страница 1: ...Operator s Manual...

Страница 2: ...play Layout 5 Display Details 5 Alarms 6 Default Alarms 6 Alarm Behavior 6 Detector Maintenance 7 Bump Testing 7 Manual Bump Test Instructions 7 Calibration 7 Manual Calibration Instructions 7 Battery...

Страница 3: ...bustible sensor poisons such as but not limited to alcohol and citrus based cleaners silicones lead compounds eg tetraethyl lead sulfur compounds phosphorus halogenated hydrocarbons and aerosols If su...

Страница 4: ...n If the battery is too low for activation the detector will display LOW BAT and then turn off again Once activated the unit will begin to display gas readings immediately Each sensor will show a chas...

Страница 5: ...ration bottle will be displayed when it is time to apply gas The battery is displayed in 3 bars as well as a percentage The percentage calculation is approximate and can be used to provide a rough est...

Страница 6: ...r s behavior under various alarm conditions Alarm Condition Audible Alarm Vibration Alarm Visual Alarm Low Slow sweep Slow vibration Slow LED Flash High Fast sweep Fast vibration Fast LED Flash TWA Sl...

Страница 7: ...on cap provided Once all sensors have gone into alarm the unit will return to normal operation in alarm and the bump due date will be reset If a bump test interval has not been programmed into the uni...

Страница 8: ...CT for either warranty replacement or replacement sensor s Battery The battery is displayed as both a percentage and a 3 bar battery icon The percentage calculation is approximate and can be used to p...

Страница 9: ...alarm limits Main Menu Date Time Set every time the instrument is communicated via IR Link or Docking Station User Message If the message is empty the detector will skip ahead If the startup message...

Страница 10: ...f test lock option specifies whether a button press can silence the alarm Maintenance Notification If maintenance notification is enabled the detector will periodically flash the maintenance LED Other...

Страница 11: ...80 days Calibration Gas When the detector is calibrated it scales the sensor readings to match the concentrations of the applied gases The calibration gas concentrations can be adjusted to match the r...

Страница 12: ...ese are organized by first in first out FIFO so the 26th event will replace the first event and so on Bump tests are differentiated from normal events when the alarm condition occurs inside of a Clip...

Страница 13: ...nty replacement or replacement sensor s BUMP DUE If the detector is displaying BUMP DUE the detector is either due for a bump test because of a scheduled test or has failed its last bump test Refer to...

Страница 14: ...ments O2 0 30 vol 0 1 vol increments Sensor Type H2S CO O2 Single plug in electrochemical cell Combustible Plug in Infrared IR or catalytic bead User Options User Message Language Low Alarm High Alarm...

Страница 15: ...ble gas detectors you can count on 15 of 16 Contact Information Gas Clip Technologies Inc 218 W Avenue F Midlothian TX 76065 Toll Free 877 525 0808 Phone 972 775 7577 Fax 972 775 2483 E mail info gasc...

Страница 16: ...forth in this warranty are conditional on 1 Proper storage installation calibration use maintenance and compliance with the user s manual instructions and any other applicable recommendations of GCT...

Отзывы: