Multi Gas Clip User’s Manual
V1.7
Portable gas detectors you can count on.
7 of 16
Detector Maintenance:
Bump Testing
The detector can be configured to keep track of regular bump testing intervals. When the
detector has exceeded the bump interval, it will display “BUMP DUE” until it has been
successfully bump tested. To bump test the unit, either insert the detector into the Clip
Dock test station, or manually apply gas from the bump test screen described below.
Manual Bump Test Instructions
To activate the manual bump test press the button twice rapidly to enter the user menu.
When the screen displays “Show Status” press the button once more.
The detector will flash between “BUMP DUE” and “APPLY GAS”. Apply gas to the unit at a
flow rate of 0.25 to 0.5LPM using the calibration cap provided. Once all sensors have gone
into alarm, the unit will return to normal operation (in alarm) and the bump due date will be
reset.
If a bump test interval has not been programmed into the unit, users may simply apply gas
to the instrument during normal operation. However this will not be recognized as a bump
test by the detector and therefore it will be recorded into the event log as an event.
Calibration
To detector can be configured to keep track of regular calibration intervals. When the
detector has exceeded the calibration interval, it will display “CAL DUE” until it has been
successfully calibrated. To calibrate the detector, either insert the detector into the Clip
Dock test station, or manually calibrate the detector as described below.
Manual Calibration Instructions
To enter calibration mode, manually press and hold the button down as in the turn off
sequence. Continue holding the button down until the CAL screen appears and finishes the
countdown.