Multi Gas Clip User’s Manual

 

V1.7 

Portable gas detectors you can count on. 

7 of 16 

Detector Maintenance: 

Bump Testing 

The detector can be configured to keep track of regular bump testing intervals. When the 
detector has exceeded the bump interval, it will display “BUMP DUE” until it has been 
successfully bump tested. To bump test the unit, either insert the detector into the Clip 
Dock test station, or manually apply gas from the bump test screen described below.  

Manual Bump Test Instructions 

To activate the manual bump test press the button twice rapidly to enter the user menu.  
When the screen displays “Show Status” press the button once more. 

 

 

 

The detector will flash between “BUMP DUE” and “APPLY GAS”. Apply gas to the unit at a 
flow rate of 0.25 to 0.5LPM using the calibration cap provided. Once all sensors have gone 
into alarm, the unit will return to normal operation (in alarm) and the bump due date will be 
reset. 

If a bump test interval has not been programmed into the unit, users may simply apply gas 
to the instrument during normal operation. However this will not be recognized as a bump 
test by the detector and therefore it will be recorded into the event log as an event. 

Calibration  

To detector can be configured to keep track of regular calibration intervals. When the 
detector has exceeded the calibration interval, it will display “CAL DUE” until it has been 
successfully calibrated. To calibrate the detector, either insert the detector into the Clip 
Dock test station, or manually calibrate the detector as described below. 

Manual Calibration Instructions  

To enter calibration mode, manually press and hold the button down as in the turn off 
sequence. Continue holding the button down until the CAL screen appears and finishes the 
countdown. 

 

Содержание Multi Gas Clip

Страница 1: ...Operator s Manual...

Страница 2: ...play Layout 5 Display Details 5 Alarms 6 Default Alarms 6 Alarm Behavior 6 Detector Maintenance 7 Bump Testing 7 Manual Bump Test Instructions 7 Calibration 7 Manual Calibration Instructions 7 Battery...

Страница 3: ...bustible sensor poisons such as but not limited to alcohol and citrus based cleaners silicones lead compounds eg tetraethyl lead sulfur compounds phosphorus halogenated hydrocarbons and aerosols If su...

Страница 4: ...n If the battery is too low for activation the detector will display LOW BAT and then turn off again Once activated the unit will begin to display gas readings immediately Each sensor will show a chas...

Страница 5: ...ration bottle will be displayed when it is time to apply gas The battery is displayed in 3 bars as well as a percentage The percentage calculation is approximate and can be used to provide a rough est...

Страница 6: ...r s behavior under various alarm conditions Alarm Condition Audible Alarm Vibration Alarm Visual Alarm Low Slow sweep Slow vibration Slow LED Flash High Fast sweep Fast vibration Fast LED Flash TWA Sl...

Страница 7: ...on cap provided Once all sensors have gone into alarm the unit will return to normal operation in alarm and the bump due date will be reset If a bump test interval has not been programmed into the uni...

Страница 8: ...CT for either warranty replacement or replacement sensor s Battery The battery is displayed as both a percentage and a 3 bar battery icon The percentage calculation is approximate and can be used to p...

Страница 9: ...alarm limits Main Menu Date Time Set every time the instrument is communicated via IR Link or Docking Station User Message If the message is empty the detector will skip ahead If the startup message...

Страница 10: ...f test lock option specifies whether a button press can silence the alarm Maintenance Notification If maintenance notification is enabled the detector will periodically flash the maintenance LED Other...

Страница 11: ...80 days Calibration Gas When the detector is calibrated it scales the sensor readings to match the concentrations of the applied gases The calibration gas concentrations can be adjusted to match the r...

Страница 12: ...ese are organized by first in first out FIFO so the 26th event will replace the first event and so on Bump tests are differentiated from normal events when the alarm condition occurs inside of a Clip...

Страница 13: ...nty replacement or replacement sensor s BUMP DUE If the detector is displaying BUMP DUE the detector is either due for a bump test because of a scheduled test or has failed its last bump test Refer to...

Страница 14: ...ments O2 0 30 vol 0 1 vol increments Sensor Type H2S CO O2 Single plug in electrochemical cell Combustible Plug in Infrared IR or catalytic bead User Options User Message Language Low Alarm High Alarm...

Страница 15: ...ble gas detectors you can count on 15 of 16 Contact Information Gas Clip Technologies Inc 218 W Avenue F Midlothian TX 76065 Toll Free 877 525 0808 Phone 972 775 7577 Fax 972 775 2483 E mail info gasc...

Страница 16: ...forth in this warranty are conditional on 1 Proper storage installation calibration use maintenance and compliance with the user s manual instructions and any other applicable recommendations of GCT...

Отзывы: