Timing
The following timings apply for PP600 V017.
A.1 Selected Continuous Mode
The maximum delay from an input changing to the output current changing is
500us.
A.2 Pulse Mode
For short delays and pulse widths the actual timings are repeatable but not
exact. Timings are dependent on the output current as described in the two
tables below.
The timings given are for channel 1 and assume a trigger input signal of 5V.
The delays are about 8us shorter for a 24V trigger input. The delays for channel
2 will be greater. All times are in microseconds.
Programmed
Actual timing
Output current = 100mA
Actual timing
Output current = 1 A
Delay
Pulse
Width
Delay
Pulse
Width
Delay
Pulse
Width
20
20
41
11
29
23
20
40
41
30
29
41
40
20
58
14
48
22
40
40
58
32
48
40
60
60
72
52
61
62
100
100
112
92
101
102
Variations in the delay and pulse widths are caused when events happen on
both channels simultaneously. Without simultaneous events, the variation is +/-
1us.