Revision 1
119
9/10/2019
5.19.2 Calibrating Analog Outputs
This procedure requires that the interior of the system is accessed. After
the calibration procedure is completed, it is necessary to ensure that the
system meets the appropriate safety classification requirements (Class 1,
Div. 2, Group B, C &D).
Each analog output must be calibrated individually. To calibrate analog outputs, a multimeter
set to measure milliamps must be connected across the analog output, and the analog output
must be powered, as the analyzer motherboard does not provide power to the analog
outputs. When the analog output is correctly powered, a green LED will glow on the
electronics board next to the AO terminal block.
To calibrate analog output 1:
a) Connect the multimeter to the AO 1 terminal block on the right side of the motherboard.
b) Press the
Calibrate AO 1
button in the GUI. In the
Analog Output 1 Calibration
box, there
are six user-adjustable fields, arranged. The Scaled Value for zero, span, and fail-safe
tell the analyzer what currents should be output to the AO when the data point being sent
to the AO has a value equal to its minimum or maximum output value or when the
analyzer is in bypass (fail-safe) mode.
c) Press the
Output Zero
button. The analog output will attempt to output 4 mA. Note the
actual mA output on a meter and enter this value into the Meter Reading Field below the
Output Zero button. Next press the Auto Cal button located below the Meter Reading
Field. This will cause the analog output to be calibrated to exactly 4 mA.
d) Once the multimeter does read this exact value, press the
Output Span
button and repeat
the process. This time, the meter should be made to read exactly the value in the scaled
value field in the span column, usually exactly 20 milliamps.
e) Calibrate the fail-safe position. The fail-safe position is the value that the analog output
will output when the analyzer is in bypass mode, and is a value outside the standard
scaled range of the analog output. For example, if the analog output is scaled to output 4
to 20 milliamps, the fail-safe value could be 3 or 21 milliamps. To calibrate the fail-safe
position, press the
Fail-Safe
button and follow the same procedures as the Zero and
span calibration.
f) Once the analog output is correctly calibrated, press the
Capture
button to write this new
configuration to the analyzer.
g) Test that the output works correctly using the
Analog Output Test
box. Enter a value
between 0 and 100 into the
Test (Percent)
field and ensure that the reading on the
multimeter is the same as the reading in the
Test mA
field. If it is not, the calibration is not
correct, and needs to be changed. If it is, the calibration is correct. Press the
Calibrate
AO 1
button again to close this field
The above process should be repeated for each analog output. Once all analog outputs are
calibrated, press the
Permanent Configuration Write to Analyzer
(
) button to permanently
write the new Analog Output configuration to the analyzer.
Содержание ProTech903W
Страница 8: ...Revision 1 8 9 10 2019...
Страница 109: ...Revision 1 109 9 10 2019 Figure 5 34 Create Modbus List Dialogue Box...
Страница 112: ...Revision 1 112 9 10 2019 Figure 5 36 Typical Modcon with Floating Point List...
Страница 141: ...Revision 1 141 9 10 2019 Figure 6 18 Status Tab...
Страница 158: ...Revision 1 158 9 10 2019 Figure 10 1 DC Power Wiring Diagram Figure 1010 2 AC Power Wiring Diagram...
Страница 159: ...Revision 1 159 9 10 2019 Figure 10 3 External Furnace Wiring Diagram Figure 10 4 AC DC Power Wiring Diagram...
Страница 160: ...Revision 1 160 9 10 2019 Figure 10 5 Solenoid Drivers Wiring Diagram Figure 10 6 Relays Wiring Diagram...
Страница 164: ...Revision 1 164 9 10 2019 Figure 10 13 Isolated RS 485 Port Figure 10 14 P17 RS232 Port to Ethernet Cable...
Страница 171: ...Revision 1 171 9 10 2019...
Страница 172: ...Revision 1 172 9 10 2019 Section 13 MSDS Information...