4.8 The Configuration Menu
When CFNG (F4) is pressed on the Main page, the display appears as shown in Figure 4-11.
The function keys lead to four menus that are described in Sections 4.8.1 through 4.8.4
mV Run State Time
6 1S1 S1 115
Configure ?
F1 F2 F3 F4
GEN DIO AIO MSEQ
Figure 4-11: The Configuration Page
4.8.1 The Gen Menu
When GEN (F1) is selected on the
Configuration
page (Figure 4-12), the
General
menu is
presented (Figure 4-16).
mV Run State Time
6 1S1 S1 115
•
TAPE
224.95
.05
•
ACK_AL OFF
•
L: mV OFF
•
L:ppm OFF
F1 F2 F3 F4
PGUP UP DOWN PGDN
Figure 4-12: The General Menu
•
TAPE
- indicates the tape length that remains. The value can be changed (e.g. when the
tape is replaced) by pressing the EDIT key on the keypad. A query will be presented and the
appropriate tape length should be entered, followed by the ENTER key.
•
ACK_AL
- This field is used to acknowledge all latched alarms. To acknowledge the latched
alarms, press the EDIT key and use the ENTER key to toggle the value to ON. Note that
any latched alarms will not unlatch if they are still in the alarm condition.
•
L mV
– This field is used to set the mV on the display to corrected mV (indicated by mV on
the display) or live uncorrected live mV (indicated by LmV on the display). This function
(On/OFF) can be edited by pressing the EDIT key followed by the ENTER key.
•
L ppm
- This field is used to set the concentration reading on the display to the last
calculated value (indicated by the absence of the L in front of the analysis units) or to the
current analysis live value (indicated by the presence of the L in front of the analysis units).
This function (On/OFF) can be edited by pressing the EDIT key followed by the ENTER key.
•
DAY -
presents the day of the week and cannot be edited.
•
TIME
- presents the present time and cannot be edited.
•
MODBUS ADD
- Presents the current Modbus unit address. This parameter can be edited
by pressing the EDIT key, typing in a new value followed by the ENTER key.
•
CURRENT
- Presents the sensor LED current in mA.
•
TEMP
- indicates current temperature of the sensor block.
•
SENSOR
- Presents the sensor firmware revision number and cannot be edited.
•
FIRM
- indicates the analyzer firmware revision number and cannot be edited.
Содержание ProTech903W
Страница 8: ...Revision 1 8 9 10 2019...
Страница 109: ...Revision 1 109 9 10 2019 Figure 5 34 Create Modbus List Dialogue Box...
Страница 112: ...Revision 1 112 9 10 2019 Figure 5 36 Typical Modcon with Floating Point List...
Страница 141: ...Revision 1 141 9 10 2019 Figure 6 18 Status Tab...
Страница 158: ...Revision 1 158 9 10 2019 Figure 10 1 DC Power Wiring Diagram Figure 1010 2 AC Power Wiring Diagram...
Страница 159: ...Revision 1 159 9 10 2019 Figure 10 3 External Furnace Wiring Diagram Figure 10 4 AC DC Power Wiring Diagram...
Страница 160: ...Revision 1 160 9 10 2019 Figure 10 5 Solenoid Drivers Wiring Diagram Figure 10 6 Relays Wiring Diagram...
Страница 164: ...Revision 1 164 9 10 2019 Figure 10 13 Isolated RS 485 Port Figure 10 14 P17 RS232 Port to Ethernet Cable...
Страница 171: ...Revision 1 171 9 10 2019...
Страница 172: ...Revision 1 172 9 10 2019 Section 13 MSDS Information...