-61-
6.6.3.6
Set Results Units
Choose the
Result (1...4)
Text box to specify the units to be displayed for results
captured in the Measurements Screen.
For Example: -
If the data captured in Result1 is Turbidity then the unit of Result1
can be given as "FTU".
6.6.3.7
Setup Data Logging
You can specify the Time Interval (Sec) with which the data needs to be stored in
buffer for each result.
Choose
Store Result1 (1…4)
text box to specify the time in seconds.
Choose
Set Buffer Size
to Specify the Buffer Size for each Result
In order to default the values Choose Set
Default Values
(i.e. 60 Seconds).
Note:
Changing Data Logging rate clears the PREVIOUS DATA of the selected
Result.
6.6.3.8
Setup Buffer Size
Choose
Result Storage Size (1.... 4)
; specify the storage size of each result.
If the data to be stored exceeds the Result storage size, Existing data is
overwritten.
Choose
Default Values
(i.e. 500)
Note:
Total maximum stored results can’t exceed 8000.
Changing the Buffer Size clears PREVIOUS DATA of ALL Results.
Содержание MONITEK MESSENGER
Страница 1: ...MESSENGER PC TRANSMITTER User s manual Release 1 2 September 15 2015...
Страница 11: ...9 Installation dimensions for Messenger PC transmitter in fiberglass enclosure...
Страница 12: ...10 Installation dimensions for Messenger PC transmitter in stainless steel enclosure...
Страница 13: ...11 1 3 2 Transmitter and Sensor Wiring System Layout and Interconnections details...
Страница 14: ...12...
Страница 15: ...13...
Страница 33: ...31 2 6 LCD Display OPTIONAL Contrast Adjust This option is not applicable to the Messenger PC Transmitter...