26
•
Repairs within warranty will be exclusively carried out by our local service partner,
SMS Elap GmbH & Co. KG, Germany, or specifically authorized companies.
6.3 Notes on maintenance
•
To preserve gears and motor for continuous untroubled operation, the UNGUATOR®
e/s should either be sent to your representative in your individual country (please
contact your UNGUATOR® dealer) or to SMS Elap GmbH & Co. KG after 20,000
preparations or after five (5) years for maintenance.
•
After-sales service and maintenance service will be billed for expenses and wearing
parts at reasonable price according to the cost estimate within the warranty period
as well.
•
Upon request an UNGUATOR® loaner may be provided for a one-time charge that can
be used while your UNGUATOR® is away for maintenance and repair.
6.4 Notes on safety
•
UNGUATOR® devices must only be connected to grounding type receptacles with
230V/50Hz(cps); 120V/60Hz(cps) or rated country specific voltage installed according
to the regulations of DIN VDE 0100.
•
UNGUATOR® devices have been designed for operation under normal room
atmospheric conditions. Recommended values: Ambient temperature between
15...30 °C/ 59...86 °F and relative air humidity less than 80 %.
•
The device should be allowed to acclimatize for approx. 30 minutes at commissioning
and/or after extended storage time in cold rooms.
•
UNGUATOR® power switch and cable must be easily accessible.
•
Do not immerse UNGUATOR® devices in water
•
UNGUATOR® devices should only be operated by authorized persons.
•
Only operate the UNGUATOR® MB with correctly attached UNGUATOR® jars or in a
plastic beaker provided for a reaction mixture.
•
Do not touch rotating parts.
•
Keep long hair, ties and cables away from rotating parts.
•
During the automatic lifting function of the UNGUATOR® e/s: always keep long hair,
parts of the body or objects away from the lifting mechanism. Immediately turn off
the power switch in an emergency or pull the power plug.
•
Always keep the air vents on the driving head/back or underside of the devices free
when using the device.
•
Using the UNGUATOR® devices not according to these operating instructions or with
line products that the manufacturer did not deliver or recommend may impair safety
and will void the warranty!
•
UNGUATOR® devices have not been designed for operation under hazardous
Gako International GmbH