- 24 -
0
Normal continuous TX
1
Special Start & Stop characters
0
Full duplex
1
Half duplex
35 (L)
Decimal Point
01 Byte values in hex
XXXXX.
02 (2 hex characters/byte) XXXX.X
03
XXX.XX
04
XX.XXX
05
X.XXXX
06
.XXXXX
34 (L)
Lockout2
0 = unlocked
1 = locked
Bit 7 6 5 4 3 2 1 0 Menu item & front panel lockout
1 Serial configuration
1
Analog output scaling
1
Alarm setpoint programming
1
Alarm setup
1
Front panel DPM reset
1
Front panel Peak & Alarm reset
1
View alarm setpoints
1
View Peak value & Tare function
33 (L)
Lockout1
0 = unlocked
1 = locked
Bit 7 6 5 4 3 2 1 0 Menu item & front panel lockout
1 Offset, Lo & Hi readings
1
Scale, Lo In, Hi In
1
Filter Setup
1
Setup, Config & Decimal Point
1
InPut Menu Item
32 (L)
Serial Cnfg2
Bit 7 6 5 4 3 2 1 0
X X X X X Binary Custom ASCII addr. 0-31
0
Continuous mode
1
Command mode
0
Alarm data not included with rdg.
1
Alarm data included with rdg.
0
No <LF> following <CR>)
1
<LF> following <CR>