Technical Instructions Related to EMC
8 - 3
Immunity information (to other than radiation/conduction)
The following is a declaration related to the electromagnetic immunity of
the equipment.
Guidance and declaration by manufacturer – Electromagnetic emission – CISPR14
The equipment is intended to be used in the electromagnetic environment designated below.
Client or user of the equipment is requested to confirm the equipment will be operated in such
environment.
Immunity test
IEC 60601
test level
Conformity level
Guidance for electromagnetic
environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
6 kV in contact.
±
8 kV in air.
±
6 kV in contact.
±
8 kV in air.
The floor must be wood, concrete or
ceramic tiles. If the floor is covered
with synthetic material, the relative
humidity must be at least 30%.
Electrical first
transient/burst
IEC 61000-4-4
±
2 kV against
power supply
line.
±
1 kV against
input/output
line.
±
2 kV against
power supply
line.
±
1 kV against
input/output
line.
The quality of main source must be
the same as standard commercial
use or hospital environment.
Surge
IEC 61000-4-5
±
1 kV in
differential mode.
±
2 kV in
common mode.
±
1 kV in
differential mode.
±
2 kV in
common mode.
The quality of main source must be
the same as standard commercial
use or hospital environment.
Voltage dip,
short-time failure
and voltage
fluctuation on
power supply input
line
IEC 61000-4-11
< 5% U
T
(drop > 95% U
T
)
for 0.5 cycle.
40% U
T
(drop of 60% U
T
)
for 5 cycles.
70% U
T
(drop of 30% U
T
)
for 25 cycles.
< 5% U
T
(drop > 95% U
T
)
for 5 seconds.
< 5% U
T
(drop > 95% U
T
)
for 0.5 cycle.
40% U
T
(drop of 60% U
T
)
for 5 cycles.
70% U
T
(drop of 30% U
T
)
for 25 cycles.
< 5% U
T
(drop > 95% U
T
)
for 5 seconds.
The quality of main source must be
the same as standard commercial
use or hospital environment.
If user of the equipment requires a
continuous operation even during
failure of main source, use of
uninterruptible power supply or
battery is recommended as power
source for the equipment.
Magnetic field
intensity of source
frequency
(50/60 Hz)
IEC 61000-4-8
3 A/m
3 A/m
The magnetic field intensity of
source frequency must be
comparable to the level
characteristics of the magnetic field
in standard commercial or hospital
environment.
Note: As U
T
, AC main source voltage has the precedence of applied test level.
Содержание FX-7302
Страница 10: ...viii ...
Страница 12: ......
Страница 60: ...Chapter 2 2 40 ...
Страница 62: ......
Страница 86: ...Chapter 4 4 18 ...
Страница 87: ...4 19 MAIN BOARD circuit diagram PCB 6956B Power connection table No 1 ...
Страница 111: ...4 43 MARK SENSOR BOARD PCB 6942 MARK SENSOR block Recoding paper 1 2 3 J1 DF4 3P 2C 5 3 4 2 1 PH1 ON2253R ...
Страница 112: ...4 44 MAGAZINE UP SENSOR BOARD PCB 6943A MAGAZINE UP SENSOR BOARD block 1 2 J1 DF4 2P 2C 1 2 3 S1 JPS1220 0401 ...
Страница 114: ...4 46 ...
Страница 116: ......
Страница 130: ......
Страница 131: ...6 1 MAIN BOARD PCB 6956B component side ...
Страница 132: ...6 2 MAIN BOARD PCB 6956B soldering side ...
Страница 133: ...6 3 OPERATION BOARD PCB 6957B component side ...
Страница 134: ...6 4 OPERATION BOARD PCB 6957B soldering side ...
Страница 135: ...6 5 ECG AMP BOARD PCB 6731D component side ...
Страница 136: ...6 6 ECG AMP BOARD PCB 6731D soldering side ...
Страница 137: ...6 7 MARK SENSER BOARD PCB 6942 component side ...
Страница 138: ...6 8 MAGAZIN UP BOARD PCB 6943A component side ...
Страница 139: ...6 9 CONNECTION BOARD PCB 7028 component side ...
Страница 140: ...6 10 ...
Страница 142: ......
Страница 151: ...7 9 Outline view 255 315 70 76 ...
Страница 152: ...7 10 ...
Страница 154: ......
Страница 160: ...Chapter 8 8 6 ...
Страница 161: ......
Страница 162: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 4R2149A 200506 F ...