Maintenance
2 - 27
Electric current calculation method
When checking a leakage current, measure the AC voltage by using a
leakage current measuring device (MD).
Convert the measured voltage to current by employing the formula below.
If the phase is switched and measurements are made, use the higher voltage
to calculate the current.
Measured voltage (V)
1000
= Measured current (mA)
Содержание FX-7302
Страница 10: ...viii ...
Страница 12: ......
Страница 60: ...Chapter 2 2 40 ...
Страница 62: ......
Страница 86: ...Chapter 4 4 18 ...
Страница 87: ...4 19 MAIN BOARD circuit diagram PCB 6956B Power connection table No 1 ...
Страница 111: ...4 43 MARK SENSOR BOARD PCB 6942 MARK SENSOR block Recoding paper 1 2 3 J1 DF4 3P 2C 5 3 4 2 1 PH1 ON2253R ...
Страница 112: ...4 44 MAGAZINE UP SENSOR BOARD PCB 6943A MAGAZINE UP SENSOR BOARD block 1 2 J1 DF4 2P 2C 1 2 3 S1 JPS1220 0401 ...
Страница 114: ...4 46 ...
Страница 116: ......
Страница 130: ......
Страница 131: ...6 1 MAIN BOARD PCB 6956B component side ...
Страница 132: ...6 2 MAIN BOARD PCB 6956B soldering side ...
Страница 133: ...6 3 OPERATION BOARD PCB 6957B component side ...
Страница 134: ...6 4 OPERATION BOARD PCB 6957B soldering side ...
Страница 135: ...6 5 ECG AMP BOARD PCB 6731D component side ...
Страница 136: ...6 6 ECG AMP BOARD PCB 6731D soldering side ...
Страница 137: ...6 7 MARK SENSER BOARD PCB 6942 component side ...
Страница 138: ...6 8 MAGAZIN UP BOARD PCB 6943A component side ...
Страница 139: ...6 9 CONNECTION BOARD PCB 7028 component side ...
Страница 140: ...6 10 ...
Страница 142: ......
Страница 151: ...7 9 Outline view 255 315 70 76 ...
Страница 152: ...7 10 ...
Страница 154: ......
Страница 160: ...Chapter 8 8 6 ...
Страница 161: ......
Страница 162: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 4R2149A 200506 F ...