C141-E064-03EN
6 - 4
When an error is detected in the self-diagnostics, the IDD terminates the SEND
DIAGNOSTIC command with the CHECK CONDITION status. At this time only when
an error is detected in the hardware function test, the LED on the front panel of the disk
drive blinks.
The INIT should issue the REQUEST SENSE command when the CHECK CONDITION
status is posted. The sense data collected by the REQUEST SENSE command indicates
the detail information of the error detected in the self-diagnostics.
The IDD status after the CHECK CONDITION status is posted differs according to the
type of the detected error.
a)
When an error is detected in the seek or write/read test, the subsequent command can
be accepted correctly. When the command other than the REQUEST SENSE and NO
OPERATION is issued from the same INIT, the error information (sense data) is
cleared.
b)
When an error is detected in the hardware function test, the IDD posts the CHECK
CONDITION status for all I/O operation request except the REQUEST SENSE
command. The error status is not cleared and the LED on the front panel continues
blinking even if the error information (sense data) is read. Only when the SCSI bus is
reset, the BUS DEVICE RESET message is issued or the power is turned off or re-
turned on, the status can be cleared. When this status is cleared, the IDD executes the
initial self-diagnostics again (see item (1)).
Refer to Chapter 3 of SCSI Logical Interface Specifications for further details of the
command specifications.
CAUTION
Data loss
When the SEND DIAGNOSTIC command terminates with the
CHECK CONDITION status, the INIT must collect the error
information using the REQUEST SENSE command. The
RECEIVE DIAGNOSTIC RESULTS command cannot read out the
error information detected in the self-diagnostics.
6.1.2
Test programs
The basic operations of the IDD itself can be checked with the self-diagnostic function.
However, to check general operations such as the host system and interface operations in a
status similar to the normal operation status, a test program that runs on the host system must
be used.
The structure and functions of the test program depend on the user system requirements.
Generally, it is recommended to provide a general input/output test program that includes
SCSI devices connected to the SCSI bus and input/output devices on other I/O ports.
Including the following test items in the test program is recommended to test the IDD
functions generally.
Содержание MAE3091LC SERIES
Страница 3: ...This page is intentionally left blank ...
Страница 11: ...This page is intentionally left blank ...
Страница 23: ...C141 E064 03EN 1 6 Figure 1 2 MAF series LP MP outer view Figure 1 3 MAE series LC outer view ...
Страница 29: ...This page is intentionally left blank ...
Страница 37: ...This page is intentionally left blank ...
Страница 55: ...C141 E064 03EN 4 2 Figure 4 1 External dimensions MAF series LC MC ...
Страница 56: ...C141 E064 03EN 4 3 Figure 4 2 External dimensions MAF series LP MP ...
Страница 57: ...C141 E064 03EN 4 4 Figure 4 3 External dimensions MAE series LC ...
Страница 58: ...C141 E064 03EN 4 5 Figure 4 4 External dimensions MAE series LP ...
Страница 59: ...C141 E064 03EN 4 6 Figure 4 5 External dimensions MAG series LC MC ...
Страница 60: ...C141 E064 03EN 4 7 Figure 4 6 External dimensions MAG series LP MP ...
Страница 72: ...C141 E064 03EN 4 19 Figure 4 22 External operator panel connector CN2 ...
Страница 83: ...This page is intentionally left blank ...
Страница 87: ...C141 E064 03EN 5 4 3 Connecting more than one IDD multi host Figure 5 1 SCSI bus connections 2 of 2 ...
Страница 97: ...C141 E064 03EN 5 14 Motor starts when power is turned on Figure 5 4 Checking the SCSI connection A 60 ...
Страница 117: ...This page is intentionally left blank ...
Страница 118: ...C141 E064 03EN B 1 APPENDIX B SETTING TERMINALS B 1 Setting Terminals This appendix describes setting terminals ...
Страница 121: ...This page is intentionally left blank ...
Страница 125: ...This page is intenitionally left blank ...
Страница 133: ......