1-3
FCR PRIMA T2 Operation Manual 897N102546
For Safe Operation
1
1.3 Safety
Prior to using this equipment, please carefully read safety precautions presented herein for you to
operate it correctly.
Laser Handling Precautions
This equipment is a Class 1 laser product (IEC 60825-1:2007).
FCR PRIMA T2 (CR-IR 392) Laser Unit Specifications
Class
: 3B
Medium
: Semiconductor laser
Wavelength
: 659 nm
Maximum output
: 68.2 mW (CW)
Maximum output under Fault condition : 130 mW (CW)
Beam divergence (parallel)
: 7°~12°
(perpendicular)
: 14°~20°
WARNING
The FCR PRIMA T2 (CR-IR 392) incorporates a Class 3B laser with maximum output of 68.2 mW.
To prevent exposure to its laser beams, observe the following precautions.
● Never open or remove the protective covers.
● Always contact a qualified service engineer immediately if you suspect there is a
malfunction.
CAUTION
Use of controls or adjustments or performance of procedures other than those specified
herein may result in hazardous radiation exposure.
Authorized Safety Standards
UL 60601-1:2003 (UL approved)
EN 60601-1:1990 + A1:1993 + A2:1995
Note, however, that UL certification does not take effect if this equipment is used mounted on board
a vehicle because it has not been applied for appropriate approval from UL-PS.
Classification
1) According to the type of protection against electrical shock
CLASS 1 EQUIPMENT
2) According to the degree of protection against electric shock
NO APPLIED PART
3) Protection against harmful ingress of water
IPXO
4) According to the degree of safety of application in the presence of a flammable anesthetics
mixture with air or with oxygen or nitrous oxide.
Equipment not suitable for use in the presence of a flammable anesthetics mixture with air or
with oxygen or nitrous oxide.
5) According to the mode of operation
CONTINUOUS OPERATION