FCR PRIMA T2 Operation Manual     897N102546

For Safe Operation

1

1.5  Precautions to Take When Using the 

FCR PRIMA T2 (CR-IR 392)

1.  Do not apply excessive force to a cassette when it is being inserted and also while it sits inside. 

Otherwise, personal injury and/or damage to the FCR PRIMA T2 (CR-IR 392) may result.

2.  When a long-view cassette is used, secure the FCR PRIMA T2 (CR-IR 392) main unit, and use a 

board to support the weight of the cassette. If excessive force is applied to the cassette inserted 

in the FCR PRIMA T2 (CR-IR 392), personal injury and/or damage to the FCR PRIMA T2 (CR-IR 

392) may result.

3.  When a long-view cassette is used, be sure to install toppling-prevention measures. 

If excessive force is applied to such a cassette being inserted in the FCR PRIMA T2 (CR-IR 

392), the FCR PRIMA T2 (CR-IR 392) may topple over and cause injury.

4.  Do not sit on the FCR PRIMA T2 (CR-IR 392) or apply your weight to it by leaning forward and 

placing your hands on the FCR PRIMA T2 (CR-IR 392) to keep the FCR PRIMA T2’s (CR-IR 

392) cassette set unit from breaking and to prevent possible injury from fragments or the like.

5.   Do not put your finger into the FCR PRIMA T2’s (CR-IR 392) cassette set unit since your finger 

may touch sharp edges inside and cut yourself.

6.  Since system sensitivity (S value) varies over time and by other factors including system 

trouble, never use it for radiation-dose and/or AEC (phototimer) setting purposes.

7.  If you use the S value as a measure of exposure dose for image-quality and exposed-dose 

control purposes or when you use FNC (Flexible Noise Control) processing, periodically 

determine whether the S value has changed noticeably in the meantime.

8.  After checking information attached to radiographic images for correctness, use such 

information for image interpretation work.

9.  Before inserting a cassette into the FCR PRIMA T2 (CR-IR 392), verify patient identification 

without fail by means of the patient’s full name, birthdate and the like.

10. Before inserting a cassette into the FCR PRIMA T2 (CR-IR 392), be sure to read the exposure menu.

11.  Before inserting a cassette into the FCR PRIMA T2 (CR-IR 392), make sure that the FCR 

PRIMA T2 (CR-IR 392) is in image-reading mode. This is because IP images are erased if the 

FCR PRIMA T2 (CR-IR 392) is in erase mode.

12. Do not subject the equipment or cassette to vibration while an image is being read. Otherwise, 

an image with unevenness or the like may be output, which will adversely affect image reading.

13. An inserted cassette should not be taken out until the cassette ejection lamp blinks (blue). 

If you try to pull out such an in-process cassette, it becomes the unremovable.

14. Do not insert an unspecified cassette into the Image Reader. If an unspecified cassette is 

inserted into the Image Reader, image reading is disabled. Be sure to use a specified cassette.

15. Do not place an unspecified IP into the cassette and place an IP into a specified cassette, 

otherwise an image that may affect image reading can be output, and it is possible that an IP, a 

cassette, or the FCR PRIMA T2 (CR-IR 392) is damaged.

16. Replace the erasure lamp when the error code prompting for replacement is displayed. 

For the replacement of the lamp, please contact your local authorized distributor. When the 

lamp fails, image erasure becomes impossible.

17.  Do not remove a cassette already subjected to exposure before image reading. It is possible 

that a read image is lost.

 

(For precautions to observe when reading images from an IP, make sure to see 

[Z.1 Precautions 

to Take Before Exposure]

 in “FCR PRIMA T2  Reference Guide” so that IPs are handled 

correctly.)

18. Use the table, cassette rack, etc. only for this equipment. Use these items following the 

instructions in the manual. Do not use them as a ladder, a seat or a chair with arm rests. Also, 

do not put unspecified goods on them. It may cause personal injury, and the items may be 

damaged.

19. Use a table or board that can withstand a load of more than 150kg (331lb) and has sufficient 

space for installing FCR PRIMA T2 (CR-IR 392) on it. 

Содержание CR-IR 392

Страница 1: ...tion Manual thoroughly before actually operating the FCR PRIMA T2 CR IR 392 system After reading this manual store it nearby the FCR PRIMA T2 CR IR 392 so that you can see it whenever necessary CR IR...

Страница 2: ...ii FCR PRIMA T2 Operation Manual 897N102546...

Страница 3: ...an those speci ed by FUJIFILM Corporation 6 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from negligence of precautions and operating methods contained in this manua...

Страница 4: ...392 Main Unit 2 2 2 2 2 Operation Panel 2 3 2 2 3 Operation Panel Display at the Time of Startup Shutdown 2 4 Chapter 3 Basic Operation 3 1 Performing Erasure Before Use IPs 3 1 3 2 System Startup Sh...

Страница 5: ...rate this equipment be sure to observe those precautions Failure to do so may cause you to subject to injuries or property damage to occur This system is classified as a medical device under EC Direct...

Страница 6: ...s functioning properly 3 2 Confirm that the ground wire has been completely connected 3 3 Make sure that all cords have been connected properly and safely 3 4 Be aware that correct diagnosis can be hi...

Страница 7: ...pect there is a malfunction CAUTION Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure Authorized Safety Standards...

Страница 8: ...electrical shock Install the equipment where no water may subject the equipment Ensure that the equipment is properly grounded Check that all the cables are properly and perfectly connected When using...

Страница 9: ...Connect the equipment into an outlet on a circuit different from that to which the other device s are connected Consult the manufacturer or field service technician for help Further information for I...

Страница 10: ...air 2 kV contact 4 kV contact 6 kV contact 2 kV air 4 kV air 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at le...

Страница 11: ...stance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occu...

Страница 12: ...tput power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 80 MHz to 800 MHz d 1 2 800 MHz to 2 5 GHz d 2 3 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73...

Страница 13: ...means of the patient s full name birthdate and the like 10 Before inserting a cassette into the FCR PRIMA T2 CR IR 392 be sure to read the exposure menu 11 Before inserting a cassette into the FCR PR...

Страница 14: ...official dealer or local presentative provided Caution Label 2 Do not connect telephone lines to LAN connector Only the IEC 60950 1 EN 60950 1 standard non shielded cables are appropriate for connecti...

Страница 15: ...Label 2 Do not connect telephone lines to LAN connector Only the IEC 60950 1 EN 60950 1 standard non shielded cables are appropriate for connection to this connector Caution Label 3 Use this terminal...

Страница 16: ...e disposed of with your household waste according to the WEEE Directive 2002 96 EC and your national law This product should be handed over to a designated collection point Improper handling of this t...

Страница 17: ...e processing the system produces radiographs that have a high diagnostic value and are easy to interpret 2 The system has a wide latitude for incident X rays so that a large amount of X ray diagnostic...

Страница 18: ...cassette that contains an exposed IP Main Power Switch Leave it turned ON ON I Main Power ON OFF O Main Power OFF Power Inlet Power inlet for this equipment External Device Connector I F Cable Used f...

Страница 19: ...ejecting the processed cassette Lights off when you remove the cassette When this lamp is on alternately with the Cassette Ready Lamp an unprocessed cassette is ejected When this lamp is on alternatel...

Страница 20: ...n completed Study by Console started cassette insertion possible ON Display status when the stand by switch is set to ON Status Power Erasing Mode Indicator Lamp Status Display LED Excessive amount of...

Страница 21: ...e degradation may result Precautions to observe before using an IP or when deleting images from an over exposed IP See Z 1 2 Precautions to Observe Before Using an IP in the Reference Guide By using t...

Страница 22: ...2 s operation panel is on If the power erasing mode indicator lamp is not lit turn ON the main power switch Power Erasing Mode Indicator Lamp Console unconnected Status when initialization has been co...

Страница 23: ...Cassette Removal Lamp should be blinking 2 On a screen from which the console can be shut down select 3 The following dialog appears Confirm that the Shutdown the PC is selected and press OK The syste...

Страница 24: ...dy Lamp is lit green on the operation panel Cassette Ready Lamp 2 Insert a cassette into the equipment in alignment with the guide on the right with the cassette s barcode window facing as illustrated...

Страница 25: ...tch on the operation panel to turn the power ON Stand by Switch The FCR PRIMA T2 starts initialization operation When the FCR PRIMA T2 is up and running the cassette ready lamp lights on the operation...

Страница 26: ...by Switch on the operation panel for four seconds Stand by Switch The status display LED changes from off line to shutdown processing When FCR PRIMA T2 shuts down the status display LED goes out See...

Страница 27: ...normally with the FCR PRIMA T2 For daily care clean the cassette set unit everyday Use the Users Checksheet when performing daily checks and maintenance Enter the following items of information into...

Страница 28: ...e equipment constantly Make a copy when you are using this check sheet Verification Details Frequency Date of verification Does the equipment start up normally Daily Is the communication possible norm...

Страница 29: ...n a process count of about 12 000 is reached Three years or when a process count of about 18 000 is reached Four years when a process count of about 24 000 is reached Five years when a process count o...

Страница 30: ...r years or when a process count of about 24 000 is reached Perform the preventive maintenance shown below Cleaning of the parts Suction cups rubber rollers Reflection plate and filter of the erasure u...

Страница 31: ...ng conditions Temperature 15 C 40 RH 30 C 80 RH Humidity 15 RH 30 C 80 RH 30 C no dew condensation Atmospheric pressure 750 hPa 1060 hPa 2 Non operating conditions Temperature 0 C 45 C no freeze conde...

Страница 32: ...A 2 Appendix A Specifications FCR PRIMA T2 Operation Manual 897N102546...

Страница 33: ...Operation Manual 897N102546 Maintenance and Inspection 1 Maintenance and Inspection Items Assigned to Specified Dealer For periodical inspection of the equipment and necessary arrangements consult our...

Страница 34: ......

Отзывы: