4-1
Care and Maintenance
4
FCR PRIMA T2 Operation Manual 897N102546
Chapter 4
Care and Maintenance
4.1 About Performing Daily Checks and
Maintenance
In order to use the FCR PRIMA T2 always in good working order, perform daily checks and
maintenance on it.
As for the daily checks, make certain that the FCR PRIMA T2 starts up normally and all equipment
that is connected to the FCR PRIMA T2 can communicate normally with the FCR PRIMA T2.
For daily care, clean the cassette set unit everyday.
Use the “Users Checksheet” when performing daily checks and maintenance. Enter the following
items of information into the “Users Checksheet”:
● Checkups:
Enter the date checks were performed.
● S value confirmation: Enter the date the S value was checked.
● Cassette set unit:
Enter the date the cassette set unit was cleaned.
4.2 Cleaning the Cassette Set Unit
Perform cleaning, taking care not to let dust into the FCR PRIMA T2.
Wipe the cassette set unit with a dry cloth.
CAUTION
Do not use organic solvents such as thinner, benzine, etc.
Содержание CR-IR 392