- 31 -
INZ-TN2ZSVS-E
4.1.3 Setting gas change time
Change gas is automatically fed when calibration is completed during measurement.
The duration of feed can be set as follows.
1) Move the cursor to <3. Setting about Gas change> using
the
key, and press the
key.
The screen switches as shown at right.
2) Select a digit using the
key, and change the value
using the
key.
Press the
key to register the setting.
The cursor moves to the most significant digit of the
setting.
The setting has now been completed.
• Gas change : After calibration is completed during measurement, sample gas is fed for a period
of time arbitrarily selected.
Description of gas change
• “30 to 300 sec” (in steps of 1 sec)
Setting range
“60 sec”
Initial value
ESC
ENT
,
ENT
,
,
Setting complete