- 28 -
INZ-TN2ZSVS-E
4.1.2 Setting zero calibration time and span concentration value
(1) Setting zero calibration time
Set the gas feed time for zero calibration.
• Zero gas feed time
: 180 sec
Initial value
1) Move the cursor to <2. Setting about Calibration>
using the
key, and press the
key.
The screen switches as shown at right.
2) Move the cursor to <1. About Zero Calibration> using
the
key, and press the
key.
3) The screen is switched and the cursor moves to the
value to be entered.
4) Select a digit using the
key, and change the value
using the
key.
Then press the
key.
The previous screen appears again.
• Zero gas feed time
: “180 to 999 sec” (in steps of 1 sec)
Setting range and contents
ESC
ESC
ENT
,
ENT
,
,
ENT
,
Setting complete