- 80 -
INZ-TN8ZSS-E
6.7.2 Analog input setting: fixed value
The description here is an example of the H
2
O input.
In such cases that the gas conditions do not change, that no external sensor is available, or that no analog in-
put terminals are left, set a fixed value. Note that if there is a large difference between the setpoint and the
actual value, the analyzer may not be able to deliver accurate measurement.
Select “H
2
O” by the
key or the
key.
Press the
key to move the cursor to
“Channel” setting.
(
)
Select “Fixed” by the
key or the
key, and press the
key.
(
)
Enter the fixed value.
Change the numeric value by the
key or
the
key, and move the digits by the
key. Press the
key to validate the set input
value.
(
)
ENT
ENT
ENT
ENT