4.1.2 Control unit
Select a location that meets the following conditions.
(1) Ambient temperature : –5 to 45°C
(2) Ambient humidity
: 90 % RH or less
(3) Power supply
: Rated voltage
: 100V to 240V AC
Rated frequency : 50Hz/60Hz
(4) A place where the instrument receives no vibration
(5) A place where the ambient air is clean
(6) A place where has enough space for maintenance work or inspection
(7) A place where induced electrical noise, such as large electrical current or spark, is low
(8) Install a breaker that meets the requirements of IEC60947-1 and IEC60947-3 in the facility in which
the analyzer is placed.
Notes:
•
The breaker should be installed close enough to the analyzer so that an operator can use it without dif-
ficulty.
•
The breaker should be identified as the one for this product.
•
The breaker must meet the rating of the analyzer.