- 60 -
INZ-TN8ZSS-E
6.4 Output hold
Before a maintenance work, be sure to carry out the setting explained in this subsection to hold the analog
output.
Select “Output Hold” from the “Menu” screen, and press the
key.
6.4.1 Output hold
Set output hold to “Last Meas.” or “Pre-set” to hold analog output. (Indication value on the “Measurement”
screen is not held. However, it is highlighted during the output hold time.)
Point the to the measurable component to
hold output by the
key or the
key,
and press the
key.
(
)
Select either “Last Meas.” or “Pre-set” by the
key or the
key. Press the
key
to validate the setting.
(
)
ENT
ENT
ENT
Output Hold
HC
ℓ
No
020
%FS
2018-04-01
12:00:00
Output Hold
HC
ℓ
Pre-set
020
%FS
2018-04-01
12:00:00