RF emissions 
CISPR 11 

Group 1 

The 

[EQUIPMENT  or  SYSTEM] 

use  RF  energy 

only for its internal function. Therefore, its 
RF  emissions  are  very  low  and  are  not 
likely  to  cause any  interference in nearby 
electronic equipment. 

RF emission 
CISPR 11 

Class B 

 

Harmonic emissions 
IEC 61000-3-2 

Not applicable 

Voltage  fluctuations/  flicker 
emissions 
IEC 61000-3-3 

Not applicable 

 

Guidance and manufacture’s declaration – electromagnetic immunity 

The 

[EQUIPMENT or SYSTEM] 

is intended for use in the electromagnetic environment specified below. The customer 

or the user of 

[EQUIPMENT or SYSTEM] 

should assure that it is used in such an environment. 

Immunity test 

IEC 60601 test level 

Compliance level 

Electromagnetic 

environment - guidance 

Electrostatic 
discharge (ESD) 
IEC 61000-4-2 

±6 kV contact 
±8 kV air 

±6 kV contact 
±8 kV air 

Floors  should  be  wood, 
concrete  or  ceramic  tile.  If 
floor 

are 

covered 

with 

synthetic 

material, 

the 

relative  humidity  should  be 
at  least  30%.  If  ESD 
interfere  with  the  operation 
of 

equipment 

,counter 

measurements 

such 

as 

wrist  strap,  grounding  shall 
be considered. 

Electrical 

fast 

transient/burst 
IEC 61000-4-4 

±2 kV for power supply 
lines 
±1  kV  for  input/output 
lines 
 

   
Not applicable 

Mains  power  quality  should 
be 

that 

of 

typical 

commercial 

or 

hospital 

environment. 

Surge 
IEC 61000-4-5 

±1 

kV 

differential 

mode.  ±2  kV  common 
mode 
 
 
 

Not applicable 

Mains  power  quality  should 
be 

that 

of 

typical 

commercial 

or 

hospital 

environment. 

Voltage  dips,  short 
interruptions 

and 

voltage 

variations 

on  power  supply 
input lines 
IEC 61000-4-11 

<5% U

T

   

(>95% dip in U

T

for 0.5 cycle 
 
40% U

T

 

(60% dip in U

T

for 5 cycles 
 
70% U

T

 

(30% dip in U

T

for 25 cycles 
 

Not applicable 

Mains  power  quality  should 
be 

that 

of 

typical 

commercial 

or 

hospital 

environment.  If  the  user  of 
the 

TL-100D

requires 

continued  operation  during 
power mains interruptions, it 
is  recommended  that  the 

TL-100D

be powered from an 

uninterruptible power supply 
or a battery. 
 

Содержание FT-B11W

Страница 1: ...manual appropriately for future reference This User Manual is suitable to the following models FT B11W FT B12W FT B13W FT B14W FT B21Y FT B22Y FT B11W V FT B12W V FTB13W V FT B14W V FT B21Y V FT B22Y...

Страница 2: ...V FT B12W V FTB13W V FT B14W V FT B21Y V FT B22Y V FT B13W UR for UART port connection FT B13W BT for Bluetooth connection Measurement Principle Oscillography Cuff Soft cuff Cuff size 290mm 72mm 5mm 1...

Страница 3: ...OFF Memory or 3 ON OFF Set Memory Life Time Machine 5 years or 10000 times Cuff 10000 times Contents 2 x 1 5V AAA alkaline batteries 3V Optional Storage case Instruction Manual Contraindications 1 Hea...

Страница 4: ...n intended operator The patient can measure transmit data and charge battery under normal circumstances and maintain the device and its accessories according to the user manual 3 If the cuff pressure...

Страница 5: ...16 If you use a cardiac pacemaker consult your physician before using the device 17 Be sure to use this device only for measuring blood pressure DO NOT use it for any other purpose 18 DO NOT use this...

Страница 6: ...d rub when clean the cuff 7 DO keep the device out of reach of children pets and insects III Name of Each Part Figure 1 Appearance LCD Display SET Button Calendar Clock Setting Mode MEMORY Button Cale...

Страница 7: ...shing down on arrow and sliding cover forward 1 Date Month Day 2 Time Hour Minute 3 Systolic Blood Pressure unit mmHg 4 Diastolic Blood Pressure unit mmHg 5 Pulse Rate unit beat minute 6 WHO Blood Pre...

Страница 8: ...avoid damage from the possibility of leaking batteries All the measurements will remain in the memory should the batteries become drained removed or replaced Figure 3 Figure 4 Clock Adjusting and Uni...

Страница 9: ...sing a 24 HOUR clock AM PM is not displayed NOTE WHEN EVERYTHING IS SET UP COMPLETELY THE MONITOR WILL AUTOMATICALLY SWITCH OFF BUT WHEN YOU PRESS ON OFF BUTTON AGAIN THE MONITOR WILL ACTIVATE Figure...

Страница 10: ...uff should be at approximately the same height as your heart TIPS FOR BLOOD PRESSURE MONITORING Relax for about 5 minutes before measurement Do not smoke or ingest caffeine at least 30 minutes prior t...

Страница 11: ...play screen See Figure9 NOTE Do not self diagnosis according to measured result Consult with your physician for further diagnosis If the device cause any discomfort during measuring process or fail to...

Страница 12: ...announced Figure 8 Figure 9 Assessing High Blood Pressure for Adults The follow standards for assessing high blood pressure without regard to age or gender have been established as a guideline accordi...

Страница 13: ...nsion 140 159 90 99 YELLOW Stage 2 Hypertension 160 179 100 109 ORANGE Stage 3 Hypertension 180 110 RED Note The graph is not exact but may be used as a guide in understanding non invasive blood press...

Страница 14: ...models FT B11W V FT B12W V FT B13W V FT B14W V FT B21Y V and FT B22Y V UART connection and Blue Tooth Function The UART connection model is FT B13W UR UART Universal Asynchronous Receiver Transmitter...

Страница 15: ...NIBP Start to measure according to the normal blood pressure monitor operation method After measuring the result will be displayed on LCD including systolic pressure diastolic pressure and pulse will...

Страница 16: ...tening which means it will be status of Listening once open As shown in Figure 2 and Figure 3 Figure 2 Figure 3 Then go to Host Slave in Write new value input below commands respectively Connect blood...

Страница 17: ...ality Reason Checkout LCD shows Low Battery icon Batteries are low Change new batteries Shows abnormal result Cuff is not tightened properly or its position is incorrect Tighten cuff correctly and ref...

Страница 18: ...tivated then stop using it Cuff inflation rate is too low or does not inflate Cuff or bladder inside the cuff leakage air Purchase a new cuff Cuff deflates too quickly Cuff has been applied too loose...

Страница 19: ...t 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequenc...

Страница 20: ...f floor are covered with synthetic material the relative humidity should be at least 30 If ESD interfere with the operation of equipment counter measurements such as wrist strap grounding shall be con...

Страница 21: ...800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field str...

Страница 22: ...rding to frequency of transmitter m 150 KHz to 80 MHz P d 167 1 80 MHz to 800 MHz P d 167 1 800 MHz to 2 5 GHz P d 333 2 0 01 0 117 0 117 0 233 0 1 0 369 0 369 0 738 1 1 167 1 167 2 333 10 3 689 3 689...

Страница 23: ...PART Symbol for Follow operating instructions IP22 Symbol for the IP classification Symbol for RF transmitters Manufacturer FUDAKANG INDUSTRIAL CO LTD Address No 8 Yinghe Road Yuanjiangyuan Managemen...

Отзывы: