STATEMENTS AND DECLARATIONS:

 

 
1. MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and 
needs to be installed and put into service according to the EMC information provided in 
the ACCOMPANYING DOCUMENTS   
 
 
2. Wireless communications equipment such as wireless home network devices, mobile 
phones,  cordless  telephones  and  their  base  stations,  walkie-talkies  can  affect  this 
equipment and should be kept at least a distance d = 3,3 m away from the equipment. 
(Note. As indicated in Table 6 of IEC 60601-1-2:2007 for ME EQUIPMENT, a typical cell 
phone with a maximum output power of 2 W yields d = 3,3 m at an IMMUNITY LEVEL of 
3 V/m) 
 

 

3. The manufacturer are available for request of circuit diagrams, component part lists, 

descriptions  ,calibration  instructions  ,or  other  information  that  will  assist  service 

personnel to repair those parts of the device 

 

4. Changes or modifications not expressly approved by the party responsible for 

compliance could void the user’s authority to operate the equipment. 

   This equipment has been tested and found to comply with the limits for a Class B digital 

device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide 
reasonable protection against harmful interference in a residential installation. This 
equipment generates, uses and can radiate radio frequency energy and, if not installed 
and used in accordance with the instructions, may cause harmful interference to radio 
communications. However, there is no guarantee that interference will not occur in a 
particular installation. 

If this equipment does cause harmful interference to radio or television reception, 
which can be determined by turning the equipment off and on, the user is encouraged 
to try to correct the interference by one or more of the following measures: 

-- Reorient or relocate the receiving antenna. 

-- Increase the separation between the equipment and receiver. 

-- Connect the equipment into an outlet on a circuit different from that to which the 

receiver is connected. 

-- Consult the dealer or an experienced radio/TV technician for help. 
5. Guidance and manufacturer’s delclaration 

 

Guidance and manufacture’s declaration – electromagnetic emission

 

The 

[EQUIPMENT or SYSTEM] i

s

 intended for use in the electromagnetic environment specified below. The customer of 

the user of the

 

 

[EQUIPMENT or SYSTEM] 

should assure that it is used in such an environment. 

Emission test 

Compliance 

Electromagnetic environment – 

guidance 

Содержание FT-B11W

Страница 1: ...manual appropriately for future reference This User Manual is suitable to the following models FT B11W FT B12W FT B13W FT B14W FT B21Y FT B22Y FT B11W V FT B12W V FTB13W V FT B14W V FT B21Y V FT B22Y...

Страница 2: ...V FT B12W V FTB13W V FT B14W V FT B21Y V FT B22Y V FT B13W UR for UART port connection FT B13W BT for Bluetooth connection Measurement Principle Oscillography Cuff Soft cuff Cuff size 290mm 72mm 5mm 1...

Страница 3: ...OFF Memory or 3 ON OFF Set Memory Life Time Machine 5 years or 10000 times Cuff 10000 times Contents 2 x 1 5V AAA alkaline batteries 3V Optional Storage case Instruction Manual Contraindications 1 Hea...

Страница 4: ...n intended operator The patient can measure transmit data and charge battery under normal circumstances and maintain the device and its accessories according to the user manual 3 If the cuff pressure...

Страница 5: ...16 If you use a cardiac pacemaker consult your physician before using the device 17 Be sure to use this device only for measuring blood pressure DO NOT use it for any other purpose 18 DO NOT use this...

Страница 6: ...d rub when clean the cuff 7 DO keep the device out of reach of children pets and insects III Name of Each Part Figure 1 Appearance LCD Display SET Button Calendar Clock Setting Mode MEMORY Button Cale...

Страница 7: ...shing down on arrow and sliding cover forward 1 Date Month Day 2 Time Hour Minute 3 Systolic Blood Pressure unit mmHg 4 Diastolic Blood Pressure unit mmHg 5 Pulse Rate unit beat minute 6 WHO Blood Pre...

Страница 8: ...avoid damage from the possibility of leaking batteries All the measurements will remain in the memory should the batteries become drained removed or replaced Figure 3 Figure 4 Clock Adjusting and Uni...

Страница 9: ...sing a 24 HOUR clock AM PM is not displayed NOTE WHEN EVERYTHING IS SET UP COMPLETELY THE MONITOR WILL AUTOMATICALLY SWITCH OFF BUT WHEN YOU PRESS ON OFF BUTTON AGAIN THE MONITOR WILL ACTIVATE Figure...

Страница 10: ...uff should be at approximately the same height as your heart TIPS FOR BLOOD PRESSURE MONITORING Relax for about 5 minutes before measurement Do not smoke or ingest caffeine at least 30 minutes prior t...

Страница 11: ...play screen See Figure9 NOTE Do not self diagnosis according to measured result Consult with your physician for further diagnosis If the device cause any discomfort during measuring process or fail to...

Страница 12: ...announced Figure 8 Figure 9 Assessing High Blood Pressure for Adults The follow standards for assessing high blood pressure without regard to age or gender have been established as a guideline accordi...

Страница 13: ...nsion 140 159 90 99 YELLOW Stage 2 Hypertension 160 179 100 109 ORANGE Stage 3 Hypertension 180 110 RED Note The graph is not exact but may be used as a guide in understanding non invasive blood press...

Страница 14: ...models FT B11W V FT B12W V FT B13W V FT B14W V FT B21Y V and FT B22Y V UART connection and Blue Tooth Function The UART connection model is FT B13W UR UART Universal Asynchronous Receiver Transmitter...

Страница 15: ...NIBP Start to measure according to the normal blood pressure monitor operation method After measuring the result will be displayed on LCD including systolic pressure diastolic pressure and pulse will...

Страница 16: ...tening which means it will be status of Listening once open As shown in Figure 2 and Figure 3 Figure 2 Figure 3 Then go to Host Slave in Write new value input below commands respectively Connect blood...

Страница 17: ...ality Reason Checkout LCD shows Low Battery icon Batteries are low Change new batteries Shows abnormal result Cuff is not tightened properly or its position is incorrect Tighten cuff correctly and ref...

Страница 18: ...tivated then stop using it Cuff inflation rate is too low or does not inflate Cuff or bladder inside the cuff leakage air Purchase a new cuff Cuff deflates too quickly Cuff has been applied too loose...

Страница 19: ...t 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequenc...

Страница 20: ...f floor are covered with synthetic material the relative humidity should be at least 30 If ESD interfere with the operation of equipment counter measurements such as wrist strap grounding shall be con...

Страница 21: ...800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field str...

Страница 22: ...rding to frequency of transmitter m 150 KHz to 80 MHz P d 167 1 80 MHz to 800 MHz P d 167 1 800 MHz to 2 5 GHz P d 333 2 0 01 0 117 0 117 0 233 0 1 0 369 0 369 0 738 1 1 167 1 167 2 333 10 3 689 3 689...

Страница 23: ...PART Symbol for Follow operating instructions IP22 Symbol for the IP classification Symbol for RF transmitters Manufacturer FUDAKANG INDUSTRIAL CO LTD Address No 8 Yinghe Road Yuanjiangyuan Managemen...

Отзывы: