Align the Sample Extractor Assembly door with the hinge brackets and
3.
reinstall the quick release pins.
Reinstall the front antechamber plate with the selector arm using the
4.
thumbscrews.
Sample Extractor Operation
Before beginning operation of the Standard Sample Extractor Assembly, ensure that
the front plate is closed and that the inert gas/vacuum break valve is closed.
To operate the Sample Extractor Assembly:
Activate the auxiliary vacuum pump to pull vacuum in the antechamber.
1.
After the vacuum gauge stabilizes at the maximum position, allow the vacuum
2.
pump to operate for an additional 15 minutes.
Open the antechamber isolation door. With locking handle in the up position
3.
(unlocked), slide the antechamber door open.
When the antechamber isolation door is fully open, lower the handle to lock
4.
the isolation door in position.
Raise handle to
unlock the door.
Antechamber
Isolation Door
Sample Extractor Assembly Door Without
Front Antechamber Cover And Selector Arm.
Содержание LYOSTAR 3
Страница 1: ...LYOPHILIZER OPERATOR S MANUAL FTS SYSTEMS LYOSTAR 3 RESEARCH DEVELOPMENT PILOT SCALE FREEZE DRYER...
Страница 2: ......
Страница 6: ...iv Rev 006 01 14 SP Scientific 2014...
Страница 10: ...Contents FTS Systems LyoStar 3 viii Rev 006 01 14 SP Scientific 2014...
Страница 18: ...Getting Started FTS Systems LyoStar 3 8 Rev 006 01 14 SP Scientific 2014...
Страница 30: ...Operating your Lyophilizer FTS Systems LyoStar 3 20 Rev 006 01 14 SP Scientific 2014...
Страница 36: ...Liquid Nitrogen Trap FTS Systems LyoStar 3 26 Rev 006 01 14 SP Scientific 2014...
Страница 48: ...Sample Extractor Assemblies FTS Systems LyoStar 3 38 Rev 006 01 14 SP Scientific 2014...
Страница 52: ...Stoppering FTS Systems LyoStar 3 42 Rev 006 01 14 SP Scientific 2014...
Страница 56: ...Shelf Latching FTS Systems LyoStar 3 46 Rev 006 01 14 SP Scientific 2014...
Страница 64: ...Praxair s ControLyo Nucleation On Demand Technology FTS Systems LyoStar 3 54 Rev 006 01 14 SP Scientific 2014...
Страница 74: ...Appendix A Sample Recipe FTS Systems LyoStar 3 64 Rev 006 01 14 SP Scientific 2014...
Страница 80: ...Appendix C Stainless Steel Cleaners FTS Systems LyoStar 3 70 Rev 006 01 14 SP Scientific 2014...
Страница 81: ......