Liquid Nitrogen Trap
FTS Systems LyoStar™ 3
22
Rev 006, 01/14
© SP Scientific 2014
Operation
Before you begin operating the Liquid Nitrogen Trap, read the Liquid Level
1.
Controller manual and familiarize all personnel with the safe handling and
storage of liquefied gases.
Ensure that all three reservoir cover latches have been closed and the
2.
reservoir cover is secure.
Fully insert the level sensor through the reservoir cover. If the sensor is not
3.
fully inserted, liquid nitrogen may overflow and spill out of the vent. Do not
obstruct the vent line, as it could result in a pressure build up.
Open the liquid nitrogen supply valve.
4.
Switch on the Liquid Level Controller and allow time for the liquid nitrogen
5.
reservoir to fill. Ensure that the reservoir cover is properly seated. If it is not,
the proximity sensor will prevent the cryogenic solenoid valve from opening.
When the liquid level in the reservoir reaches 4 inches (10.16 cm), the
controller will close the solenoid valve.
Note: The controller will activate the low-level audible alarm until liquid level rises
above the end of the sensor. It may be muted.
Level Sensor positioned correctly
Cover Safety (Proximity Switch)
Latch closed
Cover is secured
Vent line
Latch closed
Latch closed
Содержание LYOSTAR 3
Страница 1: ...LYOPHILIZER OPERATOR S MANUAL FTS SYSTEMS LYOSTAR 3 RESEARCH DEVELOPMENT PILOT SCALE FREEZE DRYER...
Страница 2: ......
Страница 6: ...iv Rev 006 01 14 SP Scientific 2014...
Страница 10: ...Contents FTS Systems LyoStar 3 viii Rev 006 01 14 SP Scientific 2014...
Страница 18: ...Getting Started FTS Systems LyoStar 3 8 Rev 006 01 14 SP Scientific 2014...
Страница 30: ...Operating your Lyophilizer FTS Systems LyoStar 3 20 Rev 006 01 14 SP Scientific 2014...
Страница 36: ...Liquid Nitrogen Trap FTS Systems LyoStar 3 26 Rev 006 01 14 SP Scientific 2014...
Страница 48: ...Sample Extractor Assemblies FTS Systems LyoStar 3 38 Rev 006 01 14 SP Scientific 2014...
Страница 52: ...Stoppering FTS Systems LyoStar 3 42 Rev 006 01 14 SP Scientific 2014...
Страница 56: ...Shelf Latching FTS Systems LyoStar 3 46 Rev 006 01 14 SP Scientific 2014...
Страница 64: ...Praxair s ControLyo Nucleation On Demand Technology FTS Systems LyoStar 3 54 Rev 006 01 14 SP Scientific 2014...
Страница 74: ...Appendix A Sample Recipe FTS Systems LyoStar 3 64 Rev 006 01 14 SP Scientific 2014...
Страница 80: ...Appendix C Stainless Steel Cleaners FTS Systems LyoStar 3 70 Rev 006 01 14 SP Scientific 2014...
Страница 81: ......