700-DTS-12 Rev. 28 - 16 Apr 2020
Page 22/28
3.2.1
Measurement Command Table(ver 20 and later)
Command
Measures
Median Mode
(
factory default
)
a
M!
,
a
MC!
,
a
M0!,
a
M0C!
a
C!
,
a
CC!
Wipe before measurement
Temperature in °C
•
Median Turbidity
•
Variance of the Mean
•
Wipe Status
Mean (average) Mode
a
M!
,
a
MC!
,
a
M0!,
a
M0C!
a
C!
,
a
CC!
No wipe before measurement
•
Temperature
•
Mean Turbidity
•
Variance of the Mean
a
M1!
,
a
MC1!
,
a
C1!
,
a
CC1!
No wipe before measurement
•
Mean
•
Variance
•
Median
•
BES
•
Minimum of last 100 readings (see section
1.6.2)
•
Maximum of last 100 readings (see section
1.6.2)
•
Temperature in °C
a
M2!
,
a
MC2!
,
a
C2!
,
a
CC2!
Wipe before measurement
•
Mean
•
Variance
•
Median
•
BES
•
Min of last 100 readings (see section 1.6.2)
•
Max of last 100 readings (see section 1.6.2)
•
Temperature in °C
•
Wipe Status*
a
M3!
,
a
MC3!
,
a
C3!
,
a
CC3!
No wipe before measurement
•
Single turbidity measurement
a
M8!
,
a
MC8!
,
a
C8!
,
a
CC8!
Wipe sensor window
•
Wipe Status*
* Wipe Status Codes
0 = Wiped
1 = Too cold (below set temperature value so wipe suppressed)
2=No wipe for any other reason other than temperature suppressed
3=Attempted wipe
3.2.2
Request Data Commands
This command requests the data generated by the preceding Measurement (
M
or
C
) command. An
a
D0!
command is always the first command sent to retrieve the data. If additional data needs to be read, then an
a
D1!
command is sent, then and
a
D2!
etc., up to
a
D9!
as required to retrieve all data.
Example Measurement Command followed by a Request Data Command: