1.8
Address Configuration
The DTS-12 is shipped with default address 0 (unless shipped as part of an integrated FTS system).
1.9
Calibration
The DTS-12 should be returned to FTS for calibration on an annual basis.
1.10
Maintenance
The only field maintenance required by the DTS-12 is a periodic check of the wiper for deterioration
or fouling with biological or other material. When proceeding to the field, the following items should
be carried in order to conduct routine maintenance:
FOR ROUTINE CLEANING
FOR CHANGING WIPER BLADE
•
Soft, lint-free cloth or wipes
•
Methyl or ethyl alcohol, or mild
soap/detergent
•
Soft toothbrush
•
Replacement wiper blade
•
Pliers
•
Isopropyl alcohol
•
Clean toothbrush (ie: different toothbrush
than the one used to clean the DTS-12)
•
Loctite 242
•
Activator 7649 (to reduce cure time prior to
returning DTS-12 to the water)
1.10.1
Routine Cleaning
If fouling is detected and the wiper blade is in otherwise good condition it may be cleaned and re-
used. Use a soft lint free cloth or wipe with Methyl or Ethyl alcohol, soaps or detergents to clean the
wiper, optic surface and body of the probe. You may also use a soft toothbrush to help remove the
material. Care must be taken not to scratch the optic face as this may affect the sensor calibration.
DO NOT
use harsh or abrasive cleansers or petroleum products.
1.10.2
Changing the Wiper Blade
If the wiper blade edge appears ragged or uneven then it should be replaced. Ensure you are
working in a clean, dry area.
4
Each DTS-12 is shipped with a spare piece of wiper material packaged in a small zip-lock bag attached to the
cable. If additional replacement parts are needed, refer to Section 1.10