Immunity test
Radiated RF
IEC 61000-4-3
IEC 60601 Test level
3 V/m
80.0 MHz to 2.5 GHz
Compliance level
3 V/m
80.0 MHz to 2.5 GHz
Electromagnetic environment -guidance
Recommended separation distance
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey, It should be less than the compliance
level in each frequency range.
80MHz to 800MHz
80MHz to 2.5GHz
4. Recommended separation distances between portable and mobile RF
communications equipment and this medical LCD monitor.
- The medical LCD monitor is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled.
The customer or the user of the monitor can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the medical LCD monitor as recommended below,
according to the maximum output power of the communications equipment.
150kHz to 80MHz
V1=3Vrms
Rated maximum
output power of
transmitter [W]
Separation distance according to frequency of transmitter[m]
80MHz to 800MHz
E1=3V/m
800MHz to 2.5GHz
E1=3V/m