9 - English
Recycling
Follow local governing ordinances and recycling plans regarding the
recycling or disposal of this equipment.
EEC Safety Compliance
This medical LCD monitor unit meets the requirements of EN-60601-1
so as to conform to the Medical Device Directive 93/42/EEC
(general safety information).
Use 120V rating 5-15P type plug only in the U.S
This medical LCD monitor complies to the above standards only when used with the
supplied medical grade power supply.
19”(FS-L190*D / FS-L190*DT) - JMW190KA1200F02(BRIDGE POWER CORP.)
24”/26”(FS-L240*D / FS-L240*DT / FS-L260*D)
JMW1150KA2400F04(BRIDGE POWER CORP.)
26”LED (FS-P260*D / FS-L260*D) - BPM150S24F10(BRIDGE POWER CORP.)
32”(FS-L320*D) -JMW1180KA2400F01(BRIDGE POWER CORP.)
Caution: Make sure the power cord is the correct type that is required in your area.
This medical LCD monitor has a universal power supply that allows operation in
either 100-120V AC or 200-240V AC voltage areas (no user adjustment is required).
Use the proper power cord with correct attachment plug type. If the power source is
120 V AC, use a power cord which is a Hospital Grade Power Cord with NEMA 5-15
style plug, labeled for 125 volts AC with UL and C-UL approvals. If the power source
is a 240 V AC supply, use the tandem (T blade) type attachment plug with ground
conductor power cord that meets the respective European country's safety
regulations.
The hospital-grade plug for medical products intended for use in Denmark has
DEMKO approval and is rated 13 amps at 250Vac. Plug is recommended for use in
medical applications and specifications are being added to the standard SB
107-2-D1. Plug mates with maker's Danish hospital-grade socket. Hospital sockets
have slightly different shaped openings allowing only the hospital plug, not the
standard Danish plug, to be inserted, to protect the ac circuit in specific medical
settings.
A ground post, located on the back of the display, may be used for the purpose of
grounding the display’s chassis. Any such ground must be installed in accordance
with applicable electrical codes. The groundpost is shown on the mechanical drawing
found on page 20~22.