1. Guidance and manufacturer's declaration - electromagnetic emissions
The medical LCD monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the medical LCD monitor should assure that it is used in such an environment.
Electromagnetic environment -guidance
The medical LCD monitor uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment
The medical LCD monitor is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes
2. Guidance and manufacturer's declaration - electromagnetic immunity
This medical LCD monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the medical LCD monitor should assure that it is used in such an environment.
Compliance level
6kV Contact
8kV air
2kV for power supply
lines
1kV for input/output
lines
Emission test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations
IEC 61000-3-3
Compliance
Group 1
Class B
D
Complies
Immunity test
Electrostatic
discharge(ESD)
IEC61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
IEC 60601 Test level
6kV Contact
8kV air
2kV for power supply
lines
1kV for input/output
lines
Electromagnetic environment -guidance
Floors should be wood,concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%
Mains power quality should be that of a
typical commercial or hospital environment.