![Froilabo BM340 Скачать руководство пользователя страница 42](http://html1.mh-extra.com/html/froilabo/bm340/bm340_user-manual_2338685042.webp)
38
10.
M
AINTENANCE AND SERVICE CONTRACT
10.1
S
AFETY RULES
Before performing any servicing on the freezer, turn it off using the 24 V On/Off button and disconnect it from the mains.
Cut off at the main fuse holder (on the electrical cabinet, accessible by dismantling the bottom right section.
Remove the rack mounting system.
Withdraw the rack and place it carefully on the floor.
10.2
U
SER MAINTENANCE
1)
General indications
Due to the intense cold, microorganisms from packaging, handling and contact can survive and remain fully virulent. All
necessary precautions must be taken when accessing freezers:
-
Outer surfaces must be cleaned regularly.
-
Gloves must be worn.
-
The door must be briefly opened.
-
Tissue samples and packaging must be handled under a laminar flow hood.
The freezer must be switched off before cleaning. A water jet must not be used for cleaning, to avoid splashing the
freezer.
2)
Outer surfaces
Prior actions
Turn off the device before cleaning (remove outlet and
put on/off button located behind the unit on 0
).