16
6.
Alarms and error codes
6.1
B
ATTERY BACK UP
In the event of power outage
, the buzzer, remote alarm and CO
2
injection (optional) functions remain active,
powered by a 24V backup battery supplied as standard. The temperature display is also maintained.
Warning:
the compressors are no longer powered and thus the freezer therefore does not cool.
6.2
A
LARM DISPLAY
Error messages are displayed on the touch screen.
If a fault occurs, the screen turns red.
Temperature is displayed.
Press the “i” button for error identification.
A pictogram illustrating the fault is displayed in the centre of the
screen.
A brief error with description and action to be taken are displayed
in English beneath the pictogram.
6.3
S
TOP THE ALARM
When the alarm start, visual and sound alarms are triggered.
When the alarm is triggered, it is possible to deactivate the sound by
pressing on the black rectangle located on the left of the screen
(encircled in red on the picture).
6.4
A
LARM
H
IGH TEMPERATURE
The high temperaure alarm is inhibated during the first decrease of the freezer and is activated when the freezer reach
storage temperature (-80°C by default).